Comparative analysis of occlusion efficiency of 0.018 detachable coils in a porcine model
Authors:
- Christopher GrilliCGChristopher Grilli
- Assaf GraifAGAssaf Graif
Open Access
- 18 November 2026
- journal article
- Published by Springer Nature in CVIR Endovascular
- Vol. 9 (1)
Abstract
Purpose: To compare the procedural occlusion efficiency of common 0.018-inch detachable peripheral fibered and non-fibered embolization coils, using commercially available coil lengths, in a porcine arterial model, evaluating differences in total implanted length, time to occlusion, fluoroscopy time, and packing density. Materials and methods: Ten paired arterial sites across three swine models were embolized using 4 mm and 6 mm detachable coils. Comparisons included Embold™ Soft vs. Concerto™ Helix (4 mm), Embold Soft vs. Ruby Soft (6 mm), and Embold Fibered vs. Ruby® Standard (6 mm). Modified Thrombolysis in Myocardial Infarction (TIMI) flow grade determined progression toward occlusion. Results: Among soft coils, 4 mm Embold Soft required fewer coils on average to achieve occlusion than Concerto (1.3 ± 0.6 vs. 3.7 ± 0.6, 95% CI of 0.9, 3.8) and achieved faster average occlusion time (3:46 ± 1:54 vs. 13:50 ± 3:36, 95% CI of 0:52, 19:17). 6 mm Embold Soft required fewer coils on average to achieve occlusion than Ruby Soft (2.0 ± 1.0 vs. 4.7 ± 2.1, 95% CI of − 4.5, 9.8) and achieved faster average occlusion time (6:43 ± 4:12 vs. 19:9 ± 11:6, 95% CI of -23:29, 48:21 min:seconds). On average, Embold Fibered required fewer coils than Ruby Standard (2.8 ± 0.5 vs. 5.3 ± 1.7, 95% CI of 0.4, 4.6), less total implanted coil length (55 ± 10 cm vs. 105 ± 34 cm, 95% CI of 8.95 cm, 91.1 cm), and shorter time to occlusion (9:16 ± 2:23 vs. 20:26 ± 6:49, 95% CI from 2:08 to 20:12 min:seconds). Conclusions: In an acute, heparinized porcine model with a single operator and TIMI-based angiographic assessments, Embold Soft achieved occlusion with fewer coil deployments and in a shorter time, on average, than Concerto and Ruby Soft. This was observed when evaluating the longest commercially available coils of each type (4 mm × 30 cm Embold Soft vs. 4 mm × 10 cm Concerto and 6 mm × 60 cm Embold Soft vs. 6 mm × 30 cm Ruby Soft). Additionally, Embold Fibered achieved occlusion with fewer coils, less coil length, and in a shorter time, on average, than Ruby Standard when evaluating identical coil lengths (6 mm × 20 cm).Funding Information
- Boston Scientific Corporation
This publication has 17 references indexed in Scilit:
- Comparison of Fibered versus Nonfibered Coils for Venous Embolization in an Ovine ModelJournal of Vascular and Interventional Radiology, 2023
- The Common but Complicated Tool: Review of Embolic Materials for the Interventional RadiologistSeminars in Interventional Radiology, 2021
- Nylon Fibered Versus Non-Fibered Embolization Coils: Comparison in a Swine ModelJournal of Vascular and Interventional Radiology, 2019
- Midterm Recanalization after Arterial Embolization Using Hydrogel-Coated Coils versus Fibered Coils in an Animal ModelJournal of Vascular and Interventional Radiology, 2019
- Experience Using the Penumbra Ruby Coil in the Peripheral Vasculature: ACE Multicenter Study Preliminary ResultsJournal of Vascular and Interventional Radiology, 2015
- In-vitro thrombogenicity assessment of polymer filament modified and native platinum embolic coilsJournal of the Neurological Sciences, 2014
- Long-term Outcomes of Coil Packing for Visceral Aneurysms: Correlation between Packing Density and Incidence of Coil Compaction or RecanalizationJournal of Vascular and Interventional Radiology, 2013
- HydroSoft coil versus HydroCoil for endovascular aneurysm occlusion study: A single center experienceEuropean Journal of Radiology, 2011
- Reperfusion of Pulmonary Arteriovenous Malformations after EmbolotherapyJournal of Vascular and Interventional Radiology, 2005
- History of FDA good laboratory practicesThe Quality Assurance Journal, 2003