Abstract
We compared the alignment of 39 total knee replacements implanted using the conventional alignment guide system with 37 implanted using a CT-based navigation system, performed by a single surgeon. The knees were evaluated using full-length weight-bearing anteroposterior radiographs, lateral radiographs and CT scans. The mean hip-knee-ankle angle, coronal femoral component angle and coronal tibial component angle were 181.8° (174.2° to 188.3°), 88.5° (84.0° to 91.8°) and 89.7° (86.3° to 95.1°), respectively for the conventional group and 180.8° (178.2° to 185.1°), 89.3° (85.8° to 92.0°) and 89.9° (88.0° to 93.0°), respectively for the navigated group. The mean sagittal femoral component angle was 85.5° (80.6° to 92.8°) for the conventional group and 89.6° (85.5° to 94.0°) for the navigated group. The mean rotational femoral and tibial component angles were −0.7° (−8.8° to 9.8°) and −3.3° (−16.8° to 5.8°) for the conventional group and −0.6° (−3.5° to 3.0°) and 0.3° (−5.3° to 7.7°) for the navigated group. The ideal angles of all alignments in the navigated group were obtained at significantly higher rates than in the conventional group. Our results demonstrated significant improvements in component positioning with a CT-based navigation system, especially with respect to rotational alignment.Keywords
This publication has 40 references indexed in Scilit:
- Errors in the identification of the transepicondylar and anteroposterior axes of the distal femur in total knee replacement using minimally-invasive and conventional approachesThe Journal of Bone and Joint Surgery. British volume, 2008
- Alignment and orientation of the components in total knee replacement with and without navigation supportThe Journal of Bone and Joint Surgery. British volume, 2007
- Computer navigation versus conventional total knee replacementThe Journal of Bone and Joint Surgery. British volume, 2007
- The Chitranjan Ranawat Award: Long-term Survivorship and Failure Modes of 1000 Cemented Condylar Total Knee ArthroplastiesClinical Orthopaedics and Related Research, 2006
- Two-year Outcomes of Computed Tomography-based and Computed Tomography Free Navigation for Total Knee ArthroplastiesClinical Orthopaedics and Related Research, 2006
- Rotational Deformity in Varus Osteoarthritis of the KneeClinical Orthopaedics and Related Research, 2005
- Importance of the lateral anatomic tibial slope as a guide to the tibial cut in total knee arthroplasty in Japanese patientsJournal of Orthopaedic Science, 2005
- Navigation Improves Accuracy of Rotational Alignment in Total Knee ArthroplastyClinical Orthopaedics and Related Research, 2004
- A Comparison of Rotational Landmarks in the Distal Femur and the Tibial ShaftClinical Orthopaedics and Related Research, 2003
- Tibial shaft axis does not always serve as a correct coronal landmark in total knee arthroplasty for varus kneesThe Journal of Arthroplasty, 2003