Comparison of Polar M600 Optical Heart Rate and ECG Heart Rate during Exercise
Authors:
- Pro StergiouPSPro Stergiou
- Tak S. FungTFTak S. Fung
- Larry KatzLKLarry Katz
- 1 December 2017
- journal article
- research article
- Published by Wolters Kluwer Health in Medicine & Science in Sports & Exercise
- Vol. 49 (12) , 2600-2607
Abstract
The purpose of this study was to evaluate the accuracy of the Polar M600 optical heart rate (OHR) sensor compared with ECG heart rate (HR) measurement during various physical activities. Thirty-six subjects participated in a continuous 76-min testing session, which included rest, cycling warm-up, cycling intervals, circuit weight training, treadmill intervals, and recovery. HR was measured using a three-lead ECG configuration and a Polar M600 Sport Watch on the left wrist. Statistical analyses included OHR percent accuracy, mean difference, mean absolute error, Bland–Altman plots, and a repeated-measures generalized estimating equation design. OHR percent accuracy was calculated as the percentage of occurrences where OHR measurement was within and including ±5 bpm from the ECG HR value. Of the four exercise phases performed, the highest OHR percent accuracy was found during cycle intervals (91.8%), and the lowest OHR percent accuracy occurred during circuit weight training (34.5%). OHR percent accuracy improved steadily within exercise transitions during cycle intervals to a maximum of 98.5% and during treadmill intervals to a maximum of 89.0%. Lags in HR calculated by the Polar M600 OHR sensor existed in comparison to ECG HR, when exercise intensity changed until steady state occurred. There was a tendency for OHR underestimation during intensity increases and overestimation during intensity decreases. No statistically significant interaction effect with device was found in this sample on the basis of sex, body mass index, V˙O2max, skin type, or wrist size. The Polar M600 was accurate during periods of steady-state cycling, walking, jogging, and running, but less accurate during some exercise intensity changes, which may be attributed to factors related to total peripheral resistance changes and pulse pressure.Keywords
This publication has 10 references indexed in Scilit:
- Variable Accuracy of Wearable Heart Rate Monitors during Aerobic ExerciseMedicine & Science in Sports & Exercise, 2017
- Accuracy of Wrist-Worn Heart Rate MonitorsJAMA Cardiology, 2017
- How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?BMJ Open Sport & Exercise Medicine, 2016
- Validation of photoplethysmography as a method to detect heart rate during rest and exerciseJournal of Medical Engineering & Technology, 2015
- How accurate is pulse rate variability as an estimate of heart rate variability?International Journal of Cardiology, 2013
- Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditionsPhysiological Measurement, 2010
- The Advantages of Wearable Green Reflected PhotoplethysmographyJournal of Medical Systems, 2010
- Photoplethysmography and its application in clinical physiological measurementPhysiological Measurement, 2007
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- THE BLOOD SUPPLY OF VARIOUS SKIN AREAS AS ESTIMATED BY THE PHOTOELECTRIC PLETHYSMOGRAPHAmerican Journal of Physiology-Legacy Content, 1938