Variable beam dose rate and DMLC IMRT to moving body anatomy
- 13 October 2008
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 35 (11) , 4837-4848
Abstract
Derivation of formulas relating leaf speeds and beam dose rates for delivering planned intensity profiles to static and moving targets in dynamic multileaf collimator (DMLC) intensity modulated radiation therapy (IMRT) is presented. The analysis of equations determining algorithms for DMLC IMRT delivery under a variable beam dose rate reveals a multitude of possible delivery strategies for a given intensity map and for any given target motion patterns. From among all equivalent delivery strategies for DMLC IMRT treatments specific subclasses of strategies can be selected to provide deliveries that are particularly suitable for clinical applications providing existing delivery devices are used. Special attention is devoted to the subclass of beam dose rate variable DMLC delivery strategies to moving body anatomy that generalize existing techniques of such deliveries in Varian DMLC irradiation methodology to static body anatomy. Few examples of deliveries from this subclass of DMLC IMRT irradiations are investigated to illustrate the principle and show practical benefits of proposed techniques.Keywords
This publication has 11 references indexed in Scilit:
- 4D DMLC leaf sequencing to minimize organ at risk dose in moving anatomyMedical Physics, 2007
- Dynamic‐MLC leaf control utilizing on‐flight intensity calculations: A robust method for real‐time IMRT delivery over moving rigid targetsMedical Physics, 2007
- Real‐time DMLC IMRT delivery for mobile and deforming targetsMedical Physics, 2005
- Synchronized delivery of DMLC intensity modulated radiation therapy for stationary and moving targetsMedical Physics, 2005
- Temporo-spatial IMRT optimization: concepts, implementation and initial resultsPhysics in Medicine & Biology, 2005
- DMLC leaf‐pair optimal control for mobile, deforming targetMedical Physics, 2005
- DMLC leaf‐pair optimal control of IMRT delivery for a moving rigid targetMedical Physics, 2004
- A method for incorporating organ motion due to breathing into 3D dose calculations in the liver: Sensitivity to variations in motionMedical Physics, 2003
- Synchronized moving aperture radiation therapy (SMART): average tumour trajectory for lung patientsPhysics in Medicine & Biology, 2003
- Characterization of a commercial multileaf collimator used for intensity modulated radiation therapyMedical Physics, 2001