Laboratory Performance Testing of Venous Cannulae During Inlet Obstruction
- 8 July 2008
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 32 (7) , 566-571
Abstract
Abstract: Venous cannulae undergo continuous improvements to achieve better and safer venous drainage. Several cannula tests have been reported, though cannula performance during inlet obstruction has never been a test criterion. In this study, five different cannulae for proximal venous drainage were tested in a mock circulation that enabled measurement of hydraulic conductance after inlet obstruction by vessel collapse. Values for hydraulic conductance ranged from 1.11 × 10–2 L/min/mm Hg for a Thin‐Flex Single Stage Venous Cannula with an open‐end lighthouse tip to 1.55 × 10–2 L/min/mm Hg for a DLP VAD Venous Cannula featuring a swirled tip profile, showing a difference that amounts to nearly 40% of the lowest conductance value. Excessive venous drainage results in potentially dangerous high‐negative venous line pressures independent of cannula design. Cannulatip design featuring swirled and grooved tip structures increases drainage capacity and enhances cannula performance during inlet obstruction.Keywords
This publication has 24 references indexed in Scilit:
- Suction Due to Left Ventricular Assist: Implications for Device Control and ManagementArtificial Organs, 2007
- An In Vitro and In Vivo Study of the Detection and Reversal of Venous Collapse During Extracorporeal Life SupportArtificial Organs, 2007
- Significant reduction in blood loss in patients undergoing minimal extracorporeal circulationTransfusion Medicine, 2006
- Mini extracorporeal circuit for coronary artery bypass grafting: initial clinical and biochemical resultsPerfusion, 2004
- Miniaturization in cardiopulmonary bypassPerfusion, 2003
- Reduction of the inflammatory response following coronary bypass grafting with total minimal extracorporeal circulationEuropean Journal of Cardio-Thoracic Surgery, 2002
- Cavitation versus Degassing: In Vitro Study of the Microbubble Phenomenon Observed During Echocardiography in Patients with Mechanical Prosthetic Cardiac ValvesEchocardiography, 2002
- Investigation and quantification of the blood trauma caused by the combined dynamic forces experienced during cardiopulmonary bypassPerfusion, 2000
- Assisted venous drainage presents the risk of undetected air microembolismThe Journal of Thoracic and Cardiovascular Surgery, 2000
- Laboratory testing of femoral venous cannulae: effect of size, position and negative pressure on flowPerfusion, 1999