On the maximum operating frequency of prosthetic heart valves
- 27 June 2018
- journal article
- research article
- Published by IOP Publishing in Biomedical Physics & Engineering Express
- Vol. 4 (4) , 047007
Abstract
A criterion is presented to evaluate the maximum pulsating flow frequency beyond which prosthetic heart valves cease to properly operate. The idea is tested with the experimental measurements obtained with an biological Edwards-Carpenter valve type model. It is found that beyond a certain characteristic frequency, the valve is no longer capable of fully operating in a cyclic manner. Furthermore, based on data recently reported by other authors, this preliminary comparison suggests that for a wide range of materials, leaflet sizes and operating conditions prosthetic heart valves might fail to properly operate for frequencies beyond 4.5 Hz (or 240 ppm).Keywords
This publication has 15 references indexed in Scilit:
- Effect of the curvature of elastic plates on the evolution of pulsatile flow fieldsJournal of Fluids and Structures, 2015
- Biomechanical properties of native and tissue engineered heart valve constructsJournal of Biomechanics, 2014
- Experimental study of a model valve with flexible leaflets in a pulsatile flowJournal of Fluid Mechanics, 2013
- Study of the Velocity and Strain Fields in the Flow Through Prosthetic Heart ValvesJournal of Biomechanical Engineering, 2011
- Valvular heart disease: the next cardiac epidemicHeart, 2010
- Prevalence of Rheumatic Heart Disease Detected by Echocardiographic ScreeningNew England Journal of Medicine, 2007
- A prospective survey of patients with valvular heart disease in Europe: The Euro Heart Survey on Valvular Heart DiseaseEuropean Heart Journal, 2003
- On Accelerated Fatigue Testing of Prosthetic Heart Valves2003
- Study of arterial blood pressure by a Windkessel-type model: influence of arterial functional propertiesComputer Methods and Programs in Biomedicine, 1999
- Biomechanics1997