Real‐time volumetric MRI thermometry of focused ultrasound ablation in vivo: a feasibility study in pig liver and kidney
- 3 December 2010
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 24 (2) , 145-153
Abstract
MR thermometry offers the possibility to precisely guide high‐intensity focused ultrasound (HIFU) for the noninvasive treatment of kidney and liver tumours. The objectives of this study were to demonstrate therapy guidance by motion‐compensated, rapid and volumetric MR temperature monitoring and to evaluate the feasibility of MR‐guided HIFU ablation in these organs. Fourteen HIFU sonications were performed in the kidney and liver of five pigs under general anaesthesia using an MR‐compatible Philips HIFU platform prototype. HIFU sonication power and duration were varied. Volumetric MR thermometry was performed continuously at 1.5 T using the proton resonance frequency shift method employing a multi‐slice, single‐shot, echo‐planar imaging sequence with an update frequency of 2.5 Hz. Motion‐related suceptibility artefacts were compensated for using multi‐baseline reference images acquired prior to sonication. At the end of the experiment, the animals were sacrificed for macroscopic and microscopic examinations of the kidney, liver and skin. The standard deviation of the temperature measured prior to heating in the sonicated area was approximately 1°C in kidney and liver, and 2.5°C near the skin. The maximum temperature rise was 30°C for a sonication of 1.2 MHz in the liver over 15 s at 300 W. The thermal dose reached the lethal threshold (240CEM43) in two of six cases in the kidney and four of eight cases in the liver, but remained below this value in skin regions in the beam path. These findings were in agreement with histological analysis. Volumetric thermometry allows real‐time monitoring of the temperature at the target location in liver and kidney, as well as in surrounding tissues. Thermal ablation was more difficult to achieve in renal than in hepatic tissue even using higher acoustic energy, probably because of a more efficient heat evacuation in the kidney by perfusion. Copyright © 2010 John Wiley & Sons, Ltd.Keywords
This publication has 26 references indexed in Scilit:
- High-intensity focused ultrasound (HIFU): effective and safe therapy for hepatocellular carcinoma adjacent to major hepatic veinsEuropean Radiology, 2008
- High-intensity focused ultrasound ablation of liver tumours: can radiological assessment predict the histological response?British Journal of Radiology, 2008
- Influence of Intrahepatic Vessels on Volume and Shape of Percutaneous Thermal Ablation ZonesInvestigative Radiology, 2008
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- Real time monitoring of radiofrequency ablation based on MR thermometry and thermal dose in the pig liver in vivoEuropean Radiology, 2007
- The clinical applications of high intensity focused ultrasound in the prostateInternational Journal of Hyperthermia, 2007
- Real‐time monitoring of radiofrequency ablation of rabbit liver by respiratory‐gated quantitative temperature MRIJournal of Magnetic Resonance Imaging, 2006
- A Review of Magnetic Resonance Imaging-Guided Focused Ultrasound Surgery of Uterine FibroidsTopics in Magnetic Resonance Imaging, 2006
- Stability of real‐time MR temperature mapping in healthy and diseased human liverJournal of Magnetic Resonance Imaging, 2004
- Magnetic resonance temperature imaging for guidance of thermotherapyJournal of Magnetic Resonance Imaging, 2000