We report on a new calculable electrical capacitance standard, which will enable realizing the farad in the SI with a combined uncertainty (k=1) of the order of one part in 10 8 . To achieve this result, we developed various original mechanical systems to position and control macroscopic elements at the nanometer scale. A measurement system allowing the in-situ control of this positioning was integrated into the standard. The electrodes were held in position in a hexapod structure, which was mechanically isolated from the standard’s carrier structure to ensure that no force is applied to change the electrodes’ positions. An additional electrode, so-called “movable guard”, was introduced into the cavity formed by the electrodes to vary the cross capacitance of the standard. The lateral position of this guard in the cavity must be controlled to within 100 nm along its entire length of travel, which was achieved by means of two piezoelectric actuators. An additional capacitive measurement device between the moving guard and the five electrodes ensures that it is correctly positioned in the cavity. We demonstrated that the system described here guarantees the correct positioning of the guard within 80 nm in the cavity, regardless of its vertical position.Keywords
This publication has 14 references indexed in Scilit:
- An Initial Reproduction of SI Capacitance Unit From a New Calculable Capacitor at NIMIEEE Transactions on Instrumentation and Measurement, 2015
- Alignment and testing of the NIST calculable capacitor2014
- Toward a determination of the fine structure constant at LNE by means of a new Thompson–Lampard calculable capacitorThis paper was presented at the International Conference on Precision Physics of Simple Atomic Systems, held at École de Physique, les Houches, France, 30 May – 4 June, 2010.Canadian Journal of Physics, 2011
- Fabrication and Measurement of the Main Electrodes of the NMIA-BIPM Calculable Cross CapacitorsIEEE Transactions on Instrumentation and Measurement, 2010
- Improvements of the Measurement Chain for the Determination of the von Klitzing Constant $R_{K}$IEEE Transactions on Instrumentation and Measurement, 2008
- Determination of the von Klitzing constantRKin terms of the BNM calculable capacitor—fifteen years of investigationsMetrologia, 2003
- New NBS Measurements of the Absolute Farad and OhmIEEE Transactions on Instrumentation and Measurement, 1974
- A Calculable Standard of CapacitanceMetrologia, 1965
- New Apparatus at the National Bureau of Standards for Absolute Capacitance MeasurementIRE Transactions on Instrumentation, 1958
- A New Theorem in Electrostatics and its Application to Calculable Standards of CapacitanceNature, 1956