Influence of electron density spatial distribution and X‐ray beam quality during CT simulation on dose calculation accuracy
Open Access
- 6 April 2011
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 12 (3) , 80-89
Abstract
Impact of the various kVp settings used during computed tomography (CT) simulation that provides data for heterogeneity corrected dose distribution calculations in patients undergoing external beam radiotherapy with either high‐energy photon or electron beams have been investigated. The change of the Hounsfield Unit (HU) values due to the influence of kVp settings and geometrical distribution of various tissue substitute materials has also been studied. The impact of various kVp settings and electron density (ED) distribution on the accuracy of dose calculation in high‐energy photon beams was found to be well within 2%. In the case of dose distributions obtained with a commercially available Monte Carlo dose calculation algorithm for electron beams, differences of more than 10% were observed for different geometrical setups and kVp settings. Dose differences for the electron beams are relatively small at shallow depths but increase with depth around lower isodose values.PACS numbers: 87.57.Q‐, 87.55.D‐Keywords
This publication has 11 references indexed in Scilit:
- Assessing the effect of electron density in photon dose calculationsMedical Physics, 2006
- Accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settingsPhysics in Medicine & Biology, 2002
- Tolerance levels for quality assurance of electron density values generated from CT in radiotherapy treatment planningPhysics in Medicine & Biology, 2002
- Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient modelsMedical Physics, 2001
- CT number variations due to different image acquisition and reconstruction parameters: a thorax phantom studyComputerized Medical Imaging and Graphics, 2000
- Dose calculations for external photon beams in radiotherapyPhysics in Medicine & Biology, 1999
- Derivation of linear attenuation coefficients from CT numbers for low-energy photonsPhysics in Medicine & Biology, 1999
- Dosimetric impact of computed tomography calibration on a commercial treatment planning system for external radiation therapyRadiotherapy and Oncology, 1998
- An electron density calibration phantom for CT-based treatment planning computersMedical Physics, 1992
- Extraction of information from CT scans at different energiesMedical Physics, 1979