In silico evaluation and optimisation of magnetic resonance elastography of the liver
Open Access
- 10 November 2021
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 66 (22) , 225005
Abstract
Objective. Magnetic resonance elastography (MRE) is widely adopted as a biomarker of liver fibrosis. However, in vivo MRE accuracy is difficult to assess. Approach. Finite element model (FEM) simulation was employed to evaluate liver MRE accuracy and inform methodological optimisation. MRE data was simulated in a 3D FEM of the human torso including the liver, and compared with spin-echo echo-planar imaging MRE acquisitions. The simulated MRE results were compared with the ground truth magnitude of the complex shear modulus (∣ G *∣) for varying: (1) ground truth liver ∣ G *∣; (2) simulated imaging resolution; (3) added noise; (4) data smoothing. Motion and strain-based signal-to-noise (SNR) metrics were evaluated on the simulated data as a means to select higher-quality voxels for preparation of acquired MRE summary statistics of ∣ G *∣. Main results. The simulated MRE accuracy for a given ground truth ∣ G *∣ was found to be a function of imaging resolution, motion-SNR and smoothing. At typical imaging resolutions, it was found that due to under-sampling of the MRE wave-field, combined with motion-related noise, the reconstructed simulated ∣ G *∣ could contain errors on the scale of the difference between liver fibrosis stages, e.g. 54% error for ground truth ∣ G *∣ = 1 kPa. Optimum imaging resolutions were identified for given ground truth ∣ G *∣ and motion-SNR levels. Significance. This study provides important knowledge on the accuracy and optimisation of liver MRE. For example, for motion-SNR ≤ 5, to distinguish between liver ∣ G *∣ of 2 and 3 kPa (i.e. early-stage liver fibrosis) it was predicted that the optimum isotropic voxel size is 4–6 mm.Keywords
Funding Information
- NIHR Nottingham Biomedical Research Centre (BRC-1215-20003)
- Medical Research Council (MC_PC_16055)
This publication has 42 references indexed in Scilit:
- Patient-Activated Three-Dimensional Multifrequency Magnetic Resonance Elastography for High-Resolution Mechanical Imaging of the Liver and SpleenRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2013
- Magnetic resonance elastography of liver: Technique, analysis, and clinical applicationsJournal of Magnetic Resonance Imaging, 2013
- Model-based elastography: a survey of approaches to the inverse elasticity problemPhysics in Medicine & Biology, 2012
- An octahedral shear strain-based measure of SNR for 3D MR elastographyPhysics in Medicine & Biology, 2011
- Magnetic resonance elastography in the liver at 3 Tesla using a second harmonic approachMagnetic Resonance in Medicine, 2009
- Algebraic Helmholtz inversion in planar magnetic resonance elastographyPhysics in Medicine & Biology, 2008
- Assessment of Hepatic Fibrosis With Magnetic Resonance ElastographyClinical Gastroenterology and Hepatology, 2007
- Anthropomorphic breast phantoms for testing elastography systemsUltrasound in Medicine & Biology, 2006
- High-resolution tensor MR elastography for breast tumour detectionPhysics in Medicine & Biology, 2000
- Magnetic Resonance Elastography by Direct Visualization of Propagating Acoustic Strain WavesScience, 1995