Speech Recognition With the Advanced Combination Encoder and Transient Emphasis Spectral Maxima Strategies in Nucleus 24 Recipients
Authors:
- 1 June 2005
- journal article
- Published by American Speech Language Hearing Association in Journal of Speech, Language, and Hearing Research
- Vol. 48 (3) , 681-701
Abstract
One of the difficulties faced by cochlear implant (CI) recipients is perception of low-intensity speech cues. A. E. Vandali (2001) has developed the transient emphasis spectral maxima (TESM) strategy to amplify short-duration, low-level sounds. The aim of the present study was to determine whether speech scores would be significantly higher with TESM than with the advanced combination encoder (ACE) strategy fitted using procedures that optimize perception of soft speech and other sounds. Eight adult recipients of the Nucleus 24 CI system participated in this study. No significant differences in scores were seen between ACE and TESM for consonant-vowel nucleus-consonant (CNC) words presented at 55 and 65 dB SPL, for sentences in noise presented at 65 dB SPL at 2 different signal-to-noise ratios, or for closed-set vowels and consonants presented at 60 dB SPL. However, perception of stop consonants within CNC words presented at the lower level (55 dB SPL) was significantly higher with TESM than ACE. In addition, percentage of information transmitted for words at 55 dB SPL was significantly higher with TESM than with ACE for manner and voicing features for consonants in the initial word position. Analysis of closed-set consonants presented at 60 dB SPL revealed percentage of information transmitted for manner was significantly higher with TESM than with ACE. These improvements with TESM were small compared with those reported by Vandali for recipients of the Nucleus 22 CI system. It appears that mapping techniques used to program speech processors and improved processing capabilities of the Nucleus 24 system contributed to soft sounds being understood almost as well with ACE as with TESM. However, half of the participants preferred TESM to ACE for use in everyday life, and all but 1 used TESM in specific listening situations. Clinically, TESM may be useful to ensure the audibility of low-intensity, short-duration acoustic cues that are important for understanding speech, for recipients who are difficult to map, or if insufficient time precludes the use of mapping techniques to increase audibility of soft sound.Keywords
This publication has 28 references indexed in Scilit:
- Recognition of Speech Presented at Soft to Loud Levels by Adult Cochlear Implant Recipients of Three Cochlear Implant SystemsEar & Hearing, 2004
- Effects of Presentation Level on Phoneme and Sentence Recognition in Quiet by Cochlear Implant ListenersEar & Hearing, 2003
- An Investigation of Input Level Range for the Nucleus 24 Cochlear Implant System: Speech Perception Performance, Program Preference, and Loudness Comfort RatingsEar & Hearing, 2003
- Effects of Stimulation Rate with the Nucleus 24 ACE Speech Coding StrategyEar & Hearing, 2002
- Adaptive Dynamic Range Optimization for Cochlear Implants: A Preliminary StudyEar & Hearing, 2002
- Psychophysics of a prototype peri-modiolar cochlear implant electrode arrayHearing Research, 2001
- Effects of the Acoustical Dynamic Range on Speech Recognition with Cochlear Implants:Efectos en el rango dinámico del reconocimiento del habla con implantes coclearesInternational Journal of Audiology, 2001
- Acoustic characteristics of American English vowelsThe Journal of the Acoustical Society of America, 1995
- Voice Fundamental Frequency as an Auditory Supplement to the Speechreading of SentencesEar & Hearing, 1988
- Preferred Method For Clinical Determination Of Pure-Tone ThresholdsJournal of Speech and Hearing Disorders, 1959