A Sensor-Less Guider Contact Force Estimation Approach for Endovascular Slender Robot-Assisted Guidance System
- 24 February 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Instrumentation and Measurement
- Vol. 74 (00189456) , 1-9
Abstract
Objective: The objective of this study is to propose and validate a contact force estimation method based on guider shape prediction, aiming at estimating the contact force between the distal end of the guider and the vascular wall. Methods: A guider prediction algorithm is developed to simulate the stable shape of different contact points within a phantom model. Subsequently, a shape-contact force model is established, and the contact force is estimated based on the parametric equation of the guider shape. A suspended force measuring platform is developed to validate the contact force estimation approach. Additionally, the influence of contact position and guider stiffness on contact force is analyzed. Results: The average accuracy of contact force estimation is 91.17%. The average relative error of the guider shape is 2.5%, with a maximum offset error of 4.6 mm. During the contact force estimation experiment, the maximum relative error of the contact force is 19.83%, and the force deviation is 1.6 × 10-2 N. For Nitinol 0.4 mm at contact point 1, the relative error of the contact force is 11.58%, with a force deviation of 3.0 × 10-3 N. Conclusion: The estimated contact force aligns well with the measured force, demonstrating the feasibility of the proposed contact force estimation method. Significance: This study establishes a sensor-less contact force estimation method for vascular interventional robot guidance systems, offering insights into improving the robot's ability to navigate narrow blood vessels. The analysis of contact force provides both theoretical and experimental foundations for the design and application of vascular interventional robot guidance systems.Keywords
Funding Information
- National Natural Science Foundation of China (52175005)
This publication has 42 references indexed in Scilit:
- Design and Development of a Continuum Robot with Switching-StiffnessSoft Robotics, 2023
- Guidewire simulation of endovascular intervention: A systematic reviewThe International Journal of Medical Robotics and Computer Assisted Surgery, 2022
- Tool-tissue forces in surgery: A systematic reviewAnnals of Medicine & Surgery, 2021
- An improved matrix‐based endovascular guidewire position simulation using fusiform ternary treeThe International Journal of Medical Robotics and Computer Assisted Surgery, 2020
- Distal-end force prediction of tendon-sheath mechanisms for flexible endoscopic surgical robots using deep learningMechanism and Machine Theory, 2019
- A novel robotic system for vascular intervention: principles, performances, and applicationsInternational Journal of Computer Assisted Radiology and Surgery, 2019
- Operating force information on-line acquisition of a novel slave manipulator for vascular interventional surgeryBiomedical Microdevices, 2018
- A cooperation of catheters and guidewires-based novel remote-controlled vascular interventional robotBiomedical Microdevices, 2018
- Impact of robotics and a suspended lead suit on physician radiation exposure during percutaneous coronary interventionCardiovascular Revascularization Medicine, 2017
- Remote‐controlled vascular interventional surgery robotThe International Journal of Medical Robotics and Computer Assisted Surgery, 2010