Y-90 PET/MR imaging optimization with a Bayesian penalized likelihood reconstruction algorithm
Authors:
- José Calatayud-JordánJCJosé Calatayud-Jordán
- Nuria Carrasco-VelaNCNuria Carrasco-Vela
- José Chimeno-HernándezJCJosé Chimeno-Hernández
- Montserrat Carles-FariñaMCMontserrat Carles-Fariña
- Consuelo Olivas-ArroyoCOConsuelo Olivas-Arroyo
- Pilar Bello-ArquésPBPilar Bello-Arqués
- Daniel Pérez-EnguixDPDaniel Pérez-Enguix
- Luis Martí-BonmatíLuis Martí-Bonmatí
- Irene Torres-EspallardoITIrene Torres-Espallardo
- 17 June 2024
- journal article
- research article
- Published by Springer Nature in Physical and Engineering Sciences in Medicine
- Vol. 47 (4) , 1397-1413
Abstract
Positron Emission Tomography (PET) imaging after \(^{90}\)Y liver radioembolization is used for both lesion identification and dosimetry. Bayesian penalized likelihood (BPL) reconstruction algorithms are an alternative to ordered subset expectation maximization (OSEM) with improved image quality and lesion detectability. The investigation of optimal parameters for \(^{90}\)Y image reconstruction of Q.Clear, a commercial BPL algorithm developed by General Electric (GE), in PET/MR is a field of interest and the subject of this study. The NEMA phantom was filled at an 8:1 sphere-to-background ratio. Acquisitions were performed on a PET/MR scanner for clinically relevant activities between 0.7 and 3.3 MBq/ml. Reconstructions with Q.Clear were performed varying the \(\beta \) penalty parameter between 20 and 6000, the acquisition time between 5 and 20 min and pixel size between 1.56 and 4.69 mm. OSEM reconstructions of 28 subsets with 2 and 4 iterations with and without Time-of-Flight (TOF) were compared to Q.Clear with \(\beta \) = 4000. Recovery coefficients (RC), their coefficient of variation (COV), background variability (BV), contrast-to-noise ratio (CNR) and residual activity in the cold insert were evaluated. Increasing \(\beta \) parameter lowered RC, COV and BV, while CNR was maximized at \(\beta \) = 4000; further increase resulted in oversmoothing. For quantification purposes, \(\beta \) = 1000–2000 could be more appropriate. Longer acquisition times resulted in larger CNR due to reduced image noise. Q.Clear reconstructions led to higher CNR than OSEM. A \(\beta \) of 4000 was obtained for optimal image quality, although lower values could be considered for quantification purposes. An optimal acquisition time of 15 min was proposed considering its clinical use.
Keywords
All Related Versions
This publication has 62 references indexed in Scilit:
- Clinical oncologic applications of PET/MRI: a new horizon.2014
- Post-radioembolization yttrium-90 PET/CT - part 1: diagnostic reportingEJNMMI Research, 2013
- Assessment of acquisition protocols for routine imaging of Y-90 using PET/CTEJNMMI Research, 2013
- Quantitative 90Y image reconstruction in PETMedical Physics, 2012
- Challenges and current methods for attenuation correction in PET/MRMagnetic Resonance Materials in Physics, Biology and Medicine, 2012
- NIH Image to ImageJ: 25 years of image analysisNature Methods, 2012
- EANM procedure guideline for the treatment of liver cancer and liver metastases with intra-arterial radioactive compoundsEuropean Journal of Nuclear Medicine and Molecular Imaging, 2011
- Feasibility of 90Y TOF PET-based dosimetry in liver metastasis therapy using SIR-SpheresEuropean Journal of Nuclear Medicine and Molecular Imaging, 2010
- Partial-Volume Effect in PET Tumor ImagingJournal of Nuclear Medicine, 2007
- Radioembolization with 90Yttrium Microspheres: A State-of-the-Art Brachytherapy Treatment for Primary and Secondary Liver MalignanciesJournal of Vascular and Interventional Radiology, 2006