Optical sectioning in fluorescence microscopy
Author:
- 22 November 2010
- journal article
- review article
- Published by Wiley in Journal of Microscopy
- Vol. 242 (2) , 111-116
Abstract
We review the origins of optical sectioning in fluorescence microscopy in terms of the structure of the illumination used to generate the fluorescence within the specimen. We note that the conventional microscope using essentially uniform illumination does not exhibit optical sectioning whereas the confocal microscope using point (many spatial frequencies) illumination does. We show that the optical sectioning strength of a confocal microscope is not optimal and discuss the advantages of using a single spatial frequency for the structure of the illumination and the detection. In this case the optical sectioning strength is shown to be up to 25% narrower than in the ideal confocal case.Keywords
This publication has 6 references indexed in Scilit:
- Image formation in structured illumination wide-field fluorescence microscopyMicron, 2008
- Real time 3D fluorescence microscopy by two beam interference illuminationOptics Communications, 1998
- Method of obtaining optical sectioning by using structured light in a conventional microscopeOptics Letters, 1997
- Handbook of Biological Confocal Microscopy1995
- Optical sectioning in confocal fluorescent microscopesJournal of Microscopy, 1989
- Properties of a Defocused Optical System*Journal of the Optical Society of America, 1969