Design and characterization of a multi-frequency bioimpedance measurement prototype
- 20 December 2012
- conference paper
- Published by IOP Publishing in Journal of Physics: Conference Series
- Vol. 407 (1) , 012011
Abstract
A multi-frequency bioimpedance measurement prototype is proposed, validated and characterized. It consists of an Improved Howland Current Source controlled by voltage, a load voltage sensing scheme through a discrete 3-opamp instrumentation amplifier, a phase and quadrature demodulation setup through analog multipliers, and digitization and processing of the signals using a digital benchtop multimeter. The electrical characterization of the measurement channel was done for resistive loads only, on four different circuits. Measurements were made on 10 frequencies, from 100 kHz to 1 MHz, with 10 load resistances, from 100 Ω to 1 kΩ, to obtain linearity, absolute error and frequency response. The best performance among the four circuits was a maximum absolute error of 5.55 %, and −1.93 % of load current variation at the worst case scenario.Keywords
This publication has 6 references indexed in Scilit:
- Low-noise current excitation sub-system for medical EITPhysiological Measurement, 2008
- A review of errors in multi-frequency EIT instrumentationPhysiological Measurement, 2007
- BIOIMPEDANCE TOMOGRAPHY (ELECTRICAL IMPEDANCE TOMOGRAPHY)Annual Review of Biomedical Engineering, 2006
- Current source design for electrical impedance tomographyPhysiological Measurement, 2003
- A comparison of modified Howland circuits as current generators with current mirror type circuitsPhysiological Measurement, 2000
- Current approaches to analogue instrumentation design in electrical impedance tomographyPhysiological Measurement, 1996