Streak camera: a multidetector for diffuse optical tomography.
Authors:
- C. Virginie ZintCZC. Virginie Zint
- Wilfried UhringWUWilfried Uhring
- Murielle TorregrossaMTMurielle Torregrossa
- Bernard CuninBCBernard Cunin
- Patrick PouletPPPatrick Poulet
- 1 June 2003
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 42 (16) , 3313-3320
Abstract
We describe an experimental setup for time-resolved diffuse optical tomography that uses a seven-channel light guide to transmit scattered light to a streak camera. This setup permits the simultaneous measurement of the time profiles of photons reemitted at different boundary sites of the objects studied. The instrument, its main specifications, and detector-specific data analysis before image reconstruction are described. The instrumentation was tested with phantoms simulating biological tissue, and the absorption and reduced scattering images that were obtained are discussed.Keywords
This publication has 16 references indexed in Scilit:
- A 32-channel time-resolved instrument for medical optical tomographyReview of Scientific Instruments, 2000
- Multichannel time-resolved optical tomographic imaging systemReview of Scientific Instruments, 1999
- Optical tomography in medical imagingInverse Problems, 1999
- Quantification in tissue near–infrared spectroscopyPhilosophical Transactions of the Royal Society B: Biological Sciences, 1997
- Direct calculation of the moments of the distribution of photon time of flight in tissue with a finite-element methodApplied Optics, 1995
- Time-gated transillumination of biological tissues and tissuelike phantomsApplied Optics, 1994
- Femtosecond transillumination optical coherence tomographyOptics Letters, 1993
- A design for a stable and reproducible phantom for use in near infra-red imaging and spectroscopyPhysics in Medicine & Biology, 1993
- Optical Time-of-Flight and Absorbance Imaging of Biologic MediaScience, 1993
- Time-resolved optical tomographyApplied Optics, 1993