This paper addresses the overlearning problem in the independent component analysis (ICA) used for the removal of muscular artifacts from electroencephalographic (EEG) records. We note that for short EEG records with high number of channels the ICA fails to separate artifact-free EEG and muscular artifacts, which has been previously attributed to the phenomenon called overlearning. We address this problem by projecting an EEG record into several subspaces with a lower dimension, and perform the ICA on each subspace separately. Due to a reduced dimension of the subspaces, the overlearning is suppressed, and muscular artifacts are better separated. Once the muscular artifacts are removed, the signals in the individual subspaces are combined to provide an artifact free EEG record. We show that for short signals and high number of EEG channels our approach outperforms the currently available ICA based algorithms for muscular artifact removal. The proposed technique can efficiently suppress ICA overlearning for short signal segments of high density EEG signals.Keywords
Funding Information
- Internal CTU Grant (SGS17/183/OHK3/3T/13)
- Ministerstvo Školství, Mládeže a Tělovýchovy (CZ.02.1.01/0.0/0.0/16_019/0000765)
This publication has 25 references indexed in Scilit:
- Potential approximation in realistic Laplacian computationClinical Neurophysiology, 2013
- Enhanced ${\mu }$ Rhythm Extraction Using Blind Source Separation and Wavelet TransformIEEE Transactions on Biomedical Engineering, 2009
- Enhanced automatic artifact detection based on independent component analysis and Renyi’s entropyNeural Networks, 2008
- Scalp electrical recording during paralysis: Quantitative evidence that EEG frequencies above 20 Hz are contaminated by EMGClinical Neurophysiology, 2007
- Enhanced detection of artifacts in EEG data using higher-order statistics and independent component analysisNeuroImage, 2007
- Recovering EEG brain signals: Artifact suppression with wavelet enhanced independent component analysisJournal of Neuroscience Methods, 2006
- Removing electroencephalographic artifacts by blind source separationPsychophysiology, 2000
- Fast and robust fixed-point algorithms for independent component analysisIEEE Transactions on Neural Networks, 1999
- A blind source separation technique using second-order statisticsIEEE Transactions on Signal Processing, 1997
- Independent component analysis, A new concept?Signal Processing, 1994