Falsely Low Direct HDL-Cholesterol Results in a Patient with Dysbetalipoproteinemia

Authors:

Abstract
Direct methods for measuring HDL-cholesterol (HDL-C) are becoming more widely used because of their labor savings and favorable performance characteristics. We investigated three methodologies for direct HDL-C measurements. All three contain an initial reagent to block measurement of cholesterol in lipoproteins other than HDL, followed by a second reagent to measure HDL-C. The performance of each of these assays has been evaluated previously ( 1)( 2)( 3)( 4)( 5)( 6)( 7)( 8). All of these direct assays appear to be relatively free of common interferences; however, triglycerides >10 g/L may affect recoveries. There are minimal published data on these new direct methods with patients who have a dyslipidemia or are receiving lipid-modifying drugs. In this report, we describe a patient with type III hyperlipidemia for whom three direct HDL-C assays yielded erroneously low results compared with traditional HDL-C methods.

This publication has 10 references indexed in Scilit: