MR Imaging and Cochlear Implants with Retained Internal Magnets: Reducing Artifacts near Highly Inhomogeneous Magnetic Fields

Authors:

Abstract
Prepared with specific strategies, the radiologist and imaging team can customize an MR imaging examination to more effectively evaluate critical anatomy of the head and neck for patients with cochlear implants, reducing both subtle and prominent artifacts created by the retained internal magnet and shortening the duration of potential patient discomfort at the site of the implant. The number of patients receiving cochlear implants and auditory brainstem implants for severe to profound sensorineural hearing loss has rapidly increased. These implants consist of an internal component implanted between the skull and the temporal scalp and an external removable speech processor unit. A small magnet within the internal component is commonly used to hold the external speech processor unit in place. Several cochlear implant models have recently received U.S. Food and Drug Administration and European Economic Area regulatory approval to allow magnetic resonance (MR) imaging examinations to be performed under certain specified conditions. The small internal magnet presents a challenge for imaging of the head and neck near the implant, creating a nonlinear magnetic field inhomogeneity and significant MR imaging artifacts. Fat-saturation failures and susceptibility artifacts severely degrade image quality. Typical artifacts at diffusion-weighted imaging and accelerated imaging are exacerbated....

This publication has 26 references indexed in Scilit: