SEMAC: Slice encoding for metal artifact correction in MRI
Authors:
Open Access
- 6 March 2009
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 62 (1) , 66-76
Abstract
Magnetic resonance imaging (MRI) near metallic implants remains an unmet need because of severe artifacts, which mainly stem from large metal‐induced field inhomogeneities. This work addresses MRI near metallic implants with an innovative imaging technique called “Slice Encoding for Metal Artifact Correction” (SEMAC). The SEMAC technique corrects metal artifacts via robust encoding of each excited slice against metal‐induced field inhomogeneities. The robust slice encoding is achieved by extending a view‐angle‐tilting (VAT) spin‐echo sequence with additional z‐phase encoding. Although the VAT compensation gradient suppresses most in‐plane distortions, the z‐phase encoding fully resolves distorted excitation profiles that cause through‐plane distortions. By positioning all spins in a region‐of‐interest to their actual spatial locations, the through‐plane distortions can be corrected by summing up the resolved spins in each voxel. The SEMAC technique does not require additional hardware and can be deployed to the large installed base of whole‐body MRI systems. The efficacy of the SEMAC technique in eliminating metal‐induced distortions with feasible scan times is validated in phantom and in vivo spine and knee studies. Magn Reson Med 62:66–76, 2009.Keywords
This publication has 22 references indexed in Scilit:
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Prepolarized magnetic resonance imaging around metal orthopedic implantsMagnetic Resonance in Medicine, 2006
- Fast spin echo sequences with very long echo trains: Design of variable refocusing flip angle schedules and generation of clinical T2 contrastMagnetic Resonance in Medicine, 2006
- Correction of MR image distortions induced by metallic objects using a 3D cubic B‐spline basis set: Application to stereotactic surgical planningMagnetic Resonance in Medicine, 2005
- Reduction of blurring in view angle tilting MRIMagnetic Resonance in Medicine, 2005
- Quantitative evaluation of metal artifact reduction techniquesJournal of Magnetic Resonance Imaging, 2004
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Nonsusceptibility artifacts due to metallic objects in MR imagingJournal of Magnetic Resonance Imaging, 1995
- Homodyne detection in magnetic resonance imagingIEEE Transactions on Medical Imaging, 1991