Validation of 3D-DXA–Derived Proximal Femur Measurements Against QCT Across International Clinical Cohorts
Preprint
Abstract
Three-dimensional dual-energy X-ray absorptiometry (3D-DXA) reconstructs proximal femur models from standard scans to estimate cortical and trabecular bone parameters. The aim of this study was to evaluate 3D-DXA against quantitative computed tomography (QCT) across independent international cohorts. The study included 537 subjects from four cohorts: an adult population from Spain, a postmenopausal female population from the United States, an osteoarthrosis population and a young population, both from Japan. Subjects underwent both 3D-DXA and QCT imaging. Accuracy was assessed using linear regression and Bland-Altman analysis to evaluate systematic and random errors. 3D-DXA parameters strongly correlated with QCT across all datasets, with correlation coefficients between 0.82 and 0.97. Random errors were consistent across cohorts and ranged between 16.55 and 19.91 mg/cm 3 for integral volumetric bone mineral density (vBMD), between 13.52 and 18.47 mg/cm 3 for trabecular vBMD, and between 9.13 and 11.37 mg/cm 2 for cortical surface bone mineral density (sBMD). Systematic errors ranged between -14.84 and 4.50 mg/cm 3 for integral vBMD, between -8.31 and 14.41 mg/cm 3 for trabecular vBMD, and between -5.58 and 3.21 mg/cm 2 for cortical sBMD. The variations in systematic errors were likely attributable to differences in QCT acquisition protocols. Overall, these results demonstrate consistent agreement between 3D-DXA and QCT across sex, age, ethnicity, geographic regions, and clinical profiles. Taken together, these findings support the use of 3D-DXA as an accurate, non-invasive, and clinically accessible technology for advanced assessment of the cortical and trabecular compartments of the proximal femur.Keywords
This publication has 27 references indexed in Scilit:
- DXA-derived three-dimensional finite element models of the femur: validation against CT-based modelsBone, 2026
- Bone Structural Parameters as Measured by 3-Dimensional Dual-Energy X-Ray Absorptiometry Are Superior in Black Women and Demonstrate Unique Associations With Prior Fracture Versus White WomenEndocrine Practice, 2024
- Global Epidemiology of Hip Fractures: Secular Trends in Incidence Rate, Post‐Fracture Treatment, and All‐Cause MortalityJournal of Bone and Mineral Research, 2023
- Validation of 3D finite element models from simulated DXA images for biofidelic simulations of sideways fall impact to the hipBone, 2021
- QCT of the femur: Comparison between QCTPro CTXA and MIAF FemurBone, 2019
- Clinical Use of Quantitative Computed Tomography (QCT) of the Hip in the Management of Osteoporosis in Adults: the 2015 ISCD Official Positions—Part IJournal of Clinical Densitometry, 2015
- Volumetric DXA (VXA): A new method to extract 3D information from multiple in vivo DXA imagesJournal of Bone and Mineral Research, 2010
- Assessment of the 3-D shape and mechanics of the proximal femur using a shape template and a bone mineral density imageBiomechanics and Modeling in Mechanobiology, 2010
- The population burden of fractures originates in women with osteopenia, not osteoporosisOsteoporosis International, 2006
- Cross-calibration of liquid and solid QCT calibration standards: Corrections to the UCSF normative dataOsteoporosis International, 1993