Procedure-Specific Validation of Artificial Vertebrae
- 13 December 2017
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 65 (8) , 1852-1858
Abstract
Objective: The development of a novel hybrid patient simulator was initiated to provide a safe training possibility for novice surgeons. Integrated artificial vertebrae should be able to realistically mimic the haptics of transpedicular Vertebroplasty instrument insertion and pedicle screw placement. Therefore, new open-celled material compositions were developed, tested and validated with reference to elderly human vertebrae. Methods: Vertebroplasty tool insertion force and pedicle screw torque measurements were performed. To validate the new bone surrogates for transpedicular tool insertion, a novel parametric model of the procedure was developed identifying three characteristic insertion parameters (weighting factors, cutting and clamping forces). Furthermore, the slope of the insertion torque was used to validate the new materials against the human vertebrae for pedicle screw placement. Results: A relative error less than 6% confirmed the suitability of the parametric model for validation. The weighting factors ($28.8 \pm 10.6$) and the clamping forces ($3.7 \pm 1.5\ N/mm$) of the human reference were met by the bone surrogate with 1.25% of blowing agent ($22.3\pm 7.6$ and $3.0\pm 0.3\ N/mm$, respectively). However, no material was able to reflect the instrument cutting forces. The slope obtained during pedicle screw placement in human vertebrae was $96.0\pm 12.4\ Nm/m$. The material composition with 1% blowing agent achieved similar results ($97.6\pm 14.3\ Nm/m$). Conclusion: Two suitable materials which deliver realistic haptics during both instrument insertions were validated. The parametric model suitably modeled the transpedicular instrument insertion. Significance: These newly developed models provide a realistic haptic feedback during transpedicular instrument insertions with the potential of cement application during surgical skill training.
Keywords
Funding Information
- Austrian Research Promotion Agency (FFG)
- line Cooperation and Innovation and Project (845436)
This publication has 22 references indexed in Scilit:
- The biomechanical effect of artificial and human bone density on stopping and stripping torque during screw insertionJournal of the Mechanical Behavior of Biomedical Materials, 2013
- Simulation SpeakJournal of Surgical Education, 2011
- Technique, challenges and indications for percutaneous pedicle screw fixationJournal of Clinical Neuroscience, 2011
- Simulador cirúrgico para treinamento de dissecção do osso temporalBrazilian Journal of Otorhinolaryngology, 2010
- Morphological comparison between adolescent and adult lumbar pedicles using computerised tomography scanningSurgical and Radiologic Anatomy, 2010
- Structural properties of fourth-generation composite femurs and tibiasJournal of Biomechanics, 2008
- The importance of haptic feedback in laparoscopic suturing training and the additive value of virtual reality simulationSurgical Endoscopy, 2007
- Mechanical properties of open-cell foam synthetic thoracic vertebraeJournal of Materials Science: Materials in Medicine, 2007
- Needle insertion into soft tissue: A surveyMedical Engineering & Physics, 2006
- Early exposure to haptic feedback enhances performance in surgical simulator training: a prospective randomized crossover study in surgical residentsSurgical Endoscopy, 2006