Long‐term beam output stability of an accelerator‐based boron neutron capture therapy system
- 18 September 2024
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 51 (12) , 9250-9263
Abstract
Background: Accelerator‐based boron neutron capture therapy (AB‐BNCT) systems are becoming commercially available and are expected to be widely used in hospitals. To ensure the safety of BNCT, establishing a quality assurance (QA) program and properly managing the stability of the system are necessary. In particular, a high level of beam output stability is required to avoid accidents because beam output is a major factor in patient dose. However, no studies have analyzed the long‐term beam output stability of AB‐BNCT systems.Purpose: This study aimed to retrospectively analyze the long‐term stability of the beam output by statistical process control (SPC) based on the QA results over 3 years.Methods: The data analyzed are the results of daily QA (DQA) and weekly QA (WQA) in an AB‐BNCT system and were taken between June 2020 and September 2023. The evaluation of the stability of the beam output was based on the reaction rate between gold and neutrons calculated using the activation foil method using a gold foil. In DQA, which can be performed in a short time, the gold foil was applied directly to the beam irradiation aperture in air. In WQA, measurements were performed at the phantom surface, 2‐cm depth, and 6‐cm depth using a dedicated water phantom. The acquired data were retrospectively analyzed by individuals and a moving range chart (I‐MR chart), exponentially weighted moving average control chart (EWMA chart), and several process capability indexes (PCIs).Results: Over 99% of the DQA I‐MR chart results were within control limits, whereas the WQA I‐MR chart results showed that 1.8%, 4.1%, and 2.0% of the measurements exceeded the control limits at the surface, 2‐cm depth, and 6‐cm depth, respectively. The variation in the reaction rate of the gold foil before and after the replacement of the target was 1.0; however, the Cpmk of DQA and the 2‐ and 6‐cm depth WQAs between August 2021 and November 2022 in treatment course 2 were 0.83, 0.77, and 0.87, respectively, which were <1.0.Conclusions: The long‐term stability of beam output was confirmed using SPC in an AB‐BNCT system. The results of the control chart revealed no significant variation or drift in the beam output, and the quantitative evaluation using PCI revealed high stability. A routine QA program will enable us to provide safe BNCT.Keywords
This publication has 36 references indexed in Scilit:
- Out‐of‐field dosimetry using a validated PHITS model and computational phantom in clinical BNCTMedical Physics, 2023
- AAPM Task Group 198 Report: An implementation guide for TG 142 quality assurance of medical acceleratorsMedical Physics, 2021
- AAPM task group 224: Comprehensive proton therapy machine quality assuranceMedical Physics, 2019
- A multi‐institutional evaluation of machine performance check system on treatment beam output and symmetry using statistical process controlJournal of Applied Clinical Medical Physics, 2019
- Present status of accelerator-based BNCT: Focus on developments in ArgentinaApplied Radiation and Isotopes, 2015
- Demonstration of a high-intensity neutron source based on a liquid-lithium target for Accelerator based Boron Neutron Capture TherapyApplied Radiation and Isotopes, 2015
- BNCT for advanced or recurrent head and neck cancerApplied Radiation and Isotopes, 2014
- Experimental verification of beam characteristics for cyclotron-based epithermal neutron source (C-BENS)Applied Radiation and Isotopes, 2011
- Effectiveness of boron neutron capture therapy for recurrent head and neck malignanciesApplied Radiation and Isotopes, 2009
- Neutron Physics1964