Study on the Variation in Mechanical Performance of Medical Balloons depending on Polymer Tube Extrusion Process Variables
Open Access
- 1 January 2026
- journal article
- Published by Korean Society of Precision Engineering in Journal of the Korean Society for Precision Engineering
- Vol. 43 (1) , 79-90
Abstract
Balloon catheters are a key technology in medical devices, essential for minimally invasive procedures. This study quantitatively analyzes how the orientation characteristics of polymer tubes, influenced by extrusion conditions, affect the mechanical properties and compliance of the final balloon—where compliance refers to the change in diameter under external pressure. Nylon 12 tubes, with a target outer diameter of 1.2 mm and an inner diameter of 1.0 mm, were extruded under six different orientation conditions by varying the screw flow rate and puller speed. The tubes were processed under identical forming conditions, allowing for a consistent evaluation of their mechanical properties. As orientation increased, elongation decreased while yield strength increased, and these trends continued in the balloon, significantly influencing compliance. To quantitatively measure orientation, we introduced the dimensionless Deborah number. We established a curve-fitted experimental model that links extrusion conditions, polymer tube properties, and balloon compliance. This model allows for the prediction of balloon performance based on extrusion-stage parameters, providing a practical framework for process optimization. Overall, this study offers an effective quantitative indicator for forecasting balloon catheter performance based on extrusion conditions and supports the systematic design of medical balloon products.Keywords
Funding Information
- Ministry of Trade, Industry and Resources (RS-2024-00442831)
This publication has 15 references indexed in Scilit:
- Biaxial reinforcements for polyethylene medical-tubes via helical convergent flowJournal of Polymer Research, 2022
- Customizable Angioplasty Balloon-Forming Machine: Towards Precision Medicine in Coronary Bifurcation Lesion InterventionsJournal of Cardiovascular Translational Research, 2022
- Nylon—A material introduction and overview for biomedical applicationsPolymers for Advanced Technologies, 2021
- The impact of upstream contraction flow on three-dimensional polymer extrudate swell from slit diesJournal of Non-Newtonian Fluid Mechanics, 2020
- Polymer Processing and Rheology2019
- Numerical Simulation of Forming Process Conditions and Wall Thickness for Balloon2017
- Rheological, Mechanical and Barrier Properties of Multilayer Nylon/Clay Nanocomposite FilmAdvances in Polymer Technology, 2011
- Analysis of row nucleated lamellar morphology of polypropylene obtained from the cast film process: Effect of melt rheology and process conditionsPolymer Engineering & Science, 2007
- Future directions in innovative minimally invasive surgeryThe Lancet, 1999
- Percutaneous transluminal coronary angioplasty in patients with prior coronary artery bypass grafting: ten years' experienceCatheterization and Cardiovascular Diagnosis, 1994