Biocompatibility properties of a new braided biodegradable urethral stent: a comparison with a biodegradable spiral and a braided metallic stent in the rabbit urethra
- 9 March 2006
- journal article
- research article
- Published by Wiley in BJU International
- Vol. 97 (4) , 856-859
Abstract
OBJECTIVE: To compare the biocompatibility properties of a new braided biodegradable self‐reinforced poly‐ l ‐lactic acid (SR‐PLLA) urethral stent to the former spiral biodegradable SR‐PLLA stent and the stainless steel stent in a rabbit model. MATERIALS AND METHODS: In all, 54 male New Zealand White rabbits were anaesthetized and stents inserted into the prostatic urethra, three of each kind for each sample time. The rabbits were killed after 1, 3, 6, 9, 12 or 15 months and light microscopy and scanning electron microscopy used to analyse the effects. RESULTS: The disintegration of the braided SR‐PLLA stent was more closely controlled than that of the spiral SR‐PLLA stent. The metallic stent induced epithelial hyperplasia and polyposis earlier than the biodegradable stents, and in these rabbits the polyposis disappeared after the disintegration process. There were no differences in the histological analyses between the biodegradable stents, whereas the metallic stents caused the strongest inflammatory reactions. CONCLUSIONS: The braided SR‐PLLA urethral stent functioned well in the rabbit urethra and clinical studies are already planned.Keywords
This publication has 10 references indexed in Scilit:
- Expansion and Fixation Properties of a New Braided Biodegradable Urethral Stent: An Experimental Study in the RabbitJournal of Urology, 2003
- A Bioabsorbable Self-Expandable, Self-Reinforced Poly-L-Lactic Acid Urethral Stent for Recurrent Urethral Strictures: Long-Term ResultsJournal of Endourology, 2002
- Viscoelastic memory and self-expansion of self-reinforced bioabsorbable stentsBiomaterials, 2002
- Interstitial Laser Coagulation and Biodegradable Self-Expandable, Self-Reinforced Poly-L-Lactic and Poly-L-Glycolic Copolymer Spiral Stent in the Treatment of Benign Prostatic EnlargementJournal of Endourology, 2002
- A double‐blind, randomized, placebo‐controlled pilot study to investigate the effects of finasteride combined with a biodegradable self‐reinforced poly l ‐lactic acid spiral stent in patients with urinary retention caused by bladder outlet obstruction from benign prostatic hyperplasiaBJU International, 2001
- A Randomised Study to Evaluate the Efficacy of a Biodegradable Stent in the Prevention of Postoperative Urinary Retention after Interstitial Laser Coagulation of the ProstateScandinavian Journal of Urology and Nephrology, 2000
- High‐energy transurethral microwave thermotherapy for large severely obstructing prostates and the use of biodegradable stents to avoid catheterization after treatmentBritish Journal of Urology, 1997
- Biodegradable Self-Reinforced Polyglycolic Acid Spiral Stent in Prevention of Postoperative Urinary Retention After Visual Laser Ablation of the Prostate-Laser ProstatectomyJournal of Urology, 1995
- Long-Term Consequences of Urethral StentsJournal of Urology, 1993
- A bioresorbable urethral stentUrological Research, 1993