Taina Isotalo
Current Organizations / Institutes
The organizations/institutes from the scholar's latest article
Previous Organizations / Institutes
The organizations / institutes from the scholar's other articles in the past ten years
Research Subjects
Taina Isotalo has a total of 16 co-authors. Here are the top co-authors by number of shared publications:
Martti Talja
Päijät‐Häme Central Hospital, Lahti, *Tampere University of Technology, and †Tampere University Hospital, Tampere, Finland
3 shared publicationsAnu Vaajanen
Päijät‐Häme Central Hospital, Lahti, *Tampere University of Technology, and †Tampere University Hospital, Tampere, Finland
2 shared publicationsMarita Laurila
Päijät‐Häme Central Hospital, Lahti, *Tampere University of Technology, and †Tampere University Hospital, Tampere, Finland
2 shared publicationsPaula M. Martikainen
Päijät‐Häme Central Hospital, Lahti, *Tampere University of Technology, and †Tampere University Hospital, Tampere, Finland
2 shared publicationsPertti Törmälä
Päijät‐Häme Central Hospital, Lahti, *Tampere University of Technology, and †Tampere University Hospital, Tampere, Finland
2 shared publicationsTeuvo L.J. Tammela
From the Department of Surgery, Päijät-Häme Central Hospital, Lahti and Institute of Biomaterials, Tampere University of Technology, Departments of Pathology and Urology, Tampere University Hospital and Medical School, University of Tampere, Tampere, Finland
2 shared publications
Latest Publications
- 9 March 2006
- Published by Wiley in BJU International
- Vol. 97 (4) , 856-859
Biocompatibility properties of a new braided biodegradable urethral stent: a comparison with a biodegradable spiral and a braided metallic stent in the rabbit urethra
OBJECTIVE: To compare the biocompatibility properties of a new braided biodegradable self‐reinforced poly‐ l ‐lactic acid (SR‐PLLA) urethral stent to the former spiral biodegradable SR‐PLLA stent and the stainless steel stent in a rabbit model. MATERIALS AND METHODS: In all, 54 male New Zealand White rabbits were anaesthetized and stents inserted into the prostatic urethra, three of each kind for each sample time. The rabbits were killed after 1, 3, 6, 9, 12 or 15 months and light microscopy and scanning electron microscopy used to analyse the effects. RESULTS: The disintegration of the braided SR‐PLLA stent was more closely controlled than that of the spiral SR‐PLLA stent. The metallic stent induced epithelial hyperplasia and polyposis earlier than the biodegradable stents, and in these rabbits the polyposis disappeared after the disintegration process. There were no differences in the histological analyses between the biodegradable stents, whereas the metallic stents caused the strongest inflammatory reactions. CONCLUSIONS: The braided SR‐PLLA urethral stent functioned well in the rabbit urethra and clinical studies are already planned.
- 9 March 2006
- Published by Wiley in BJU International
- Vol. 97 (4) , 856-859
Biocompatibility properties of a new braided biodegradable urethral stent: a comparison with a biodegradable spiral and a braided metallic stent in the rabbit urethra
OBJECTIVE: To compare the biocompatibility properties of a new braided biodegradable self‐reinforced poly‐ l ‐lactic acid (SR‐PLLA) urethral stent to the former spiral biodegradable SR‐PLLA stent and the stainless steel stent in a rabbit model. MATERIALS AND METHODS: In all, 54 male New Zealand White rabbits were anaesthetized and stents inserted into the prostatic urethra, three of each kind for each sample time. The rabbits were killed after 1, 3, 6, 9, 12 or 15 months and light microscopy and scanning electron microscopy used to analyse the effects. RESULTS: The disintegration of the braided SR‐PLLA stent was more closely controlled than that of the spiral SR‐PLLA stent. The metallic stent induced epithelial hyperplasia and polyposis earlier than the biodegradable stents, and in these rabbits the polyposis disappeared after the disintegration process. There were no differences in the histological analyses between the biodegradable stents, whereas the metallic stents caused the strongest inflammatory reactions. CONCLUSIONS: The braided SR‐PLLA urethral stent functioned well in the rabbit urethra and clinical studies are already planned.
- 1 December 2005
- Published by Wolters Kluwer Health in Journal of Urology
- Vol. 174 (6) , 2401-2404
BIOCOMPATIBILITY AND IMPLANTATION PROPERTIES OF 2 DIFFERENTLY BRAIDED, BIODEGRADABLE, SELF-REINFORCED POLYLACTIC ACID URETHRAL STENTS: AN EXPERIMENTAL STUDY IN THE RABBIT
Biodegradable urethral stents have been in clinical use for more than 10 years. To solve the problems connected with the helical spiral configuration of the stents used to date we developed a new tubular mesh configuration and evaluated the biocompatibility properties and degradation time of 2 differently braided stents in the rabbit urethra. The biodegradable, self-expanding stents were made of self-reinforced polylactic acid polymer blended with BaSO4 (Alfa Chem, Kings Point, New York). Two braiding patterns, namely a diamond 1/1 and a regular 2/2 + 1 (Prodesco, Perkasie, Pennsylvania), were used to produce a tubular mesh configuration. Stainless steel stents with 1/1 braiding served as controls. The stents were inserted into the posterior urethra of 36 male rabbits. The animals were sacrificed after 1 week, 1 month, 6 months or 12 months. Light microscopy and scanning electron microscopy analyses were done. Tissue reactions to operative trauma were seen in all specimens at week 1. The changes gradually abated in the biodegradable stent groups, whereas chronic inflammatory changes and fibrosis were increasingly seen with metallic stents after 6 months. Epithelial hyperplasia increased with time for all stent types and materials. As expected, stent fragmentation started at 6 months. Biodegradable polymers are suitable materials for braided urethral stents. However, the braided configuration of the stent with a decreased mass of material does not prevent the development of epithelial hyperplasia. The biodegradable, self-expanding, braided stents functioned well in the rabbit urethra and are suitable for clinical studies.
- 1 December 2005
- Published by Wolters Kluwer Health in Journal of Urology
- Vol. 174 (6) , 2401-2404
BIOCOMPATIBILITY AND IMPLANTATION PROPERTIES OF 2 DIFFERENTLY BRAIDED, BIODEGRADABLE, SELF-REINFORCED POLYLACTIC ACID URETHRAL STENTS: AN EXPERIMENTAL STUDY IN THE RABBIT
Biodegradable urethral stents have been in clinical use for more than 10 years. To solve the problems connected with the helical spiral configuration of the stents used to date we developed a new tubular mesh configuration and evaluated the biocompatibility properties and degradation time of 2 differently braided stents in the rabbit urethra. The biodegradable, self-expanding stents were made of self-reinforced polylactic acid polymer blended with BaSO4 (Alfa Chem, Kings Point, New York). Two braiding patterns, namely a diamond 1/1 and a regular 2/2 + 1 (Prodesco, Perkasie, Pennsylvania), were used to produce a tubular mesh configuration. Stainless steel stents with 1/1 braiding served as controls. The stents were inserted into the posterior urethra of 36 male rabbits. The animals were sacrificed after 1 week, 1 month, 6 months or 12 months. Light microscopy and scanning electron microscopy analyses were done. Tissue reactions to operative trauma were seen in all specimens at week 1. The changes gradually abated in the biodegradable stent groups, whereas chronic inflammatory changes and fibrosis were increasingly seen with metallic stents after 6 months. Epithelial hyperplasia increased with time for all stent types and materials. As expected, stent fragmentation started at 6 months. Biodegradable polymers are suitable materials for braided urethral stents. However, the braided configuration of the stent with a decreased mass of material does not prevent the development of epithelial hyperplasia. The biodegradable, self-expanding, braided stents functioned well in the rabbit urethra and are suitable for clinical studies.
- 1 December 2002
- Published by SAGE Publications in Journal of Endourology
- Vol. 16 (10) , 759-762
A Bioabsorbable Self-Expandable, Self-Reinforced Poly-L-Lactic Acid Urethral Stent for Recurrent Urethral Strictures: Long-Term Results
Biodegradable urethral stents have been used clinically for some years mainly in order to prevent postoperative urinary retention after minimally invasive thermotreatment of benign prostatic hyperplasia. We previously reported a 15-month-follow-up pilot study of a bioabsorbable self-expandable, self-reinforced poly-L-lactic acid (SR-PLLA) urethral stent in combination with optical urethrotomy in the treatment of recurrent urethral strictures. The present aim was to evaluate the long-term results of this new treatment modality. A series of 22 patients with severe urethral strictures (mean number of earlier urethrotomies 2.5) were involved. An SR-PLLA spiral stent with a bioabsorption time of 10 to 12 months was inserted into the urethra at the stricture site immediately after optical urethrotomy. Uroflowmetry, measurement of postvoiding residual urine, urine culture, and urethroscopy were performed preoperatively and at 1 and 3 months and then every 3 months up to 12 months, with subsequent follow-up visits every 6 months. The mean follow-up was 46 months. The treatment was successful in 8 of the 22 patients (36%). Six recurrences were found within the stented area and 15 outside. Patients with failure have since been treated with several urethrotomies and repeat self-dilatations or free skin urethroplasties. The results of the use of a bioabsorbable SR-PLLA urethral stent for the treatment of recurrent strictures were encouraging. Without this additional therapy, the recurrence rate of strictures might have been much higher. The main problem was sudden collapse of the stent, possibly induced by outer compression. A new generation of bioabsorbable stents is already under development.
- 1 December 2002
- Published by SAGE Publications in Journal of Endourology
- Vol. 16 (10) , 759-762
A Bioabsorbable Self-Expandable, Self-Reinforced Poly-L-Lactic Acid Urethral Stent for Recurrent Urethral Strictures: Long-Term Results
Biodegradable urethral stents have been used clinically for some years mainly in order to prevent postoperative urinary retention after minimally invasive thermotreatment of benign prostatic hyperplasia. We previously reported a 15-month-follow-up pilot study of a bioabsorbable self-expandable, self-reinforced poly-L-lactic acid (SR-PLLA) urethral stent in combination with optical urethrotomy in the treatment of recurrent urethral strictures. The present aim was to evaluate the long-term results of this new treatment modality. A series of 22 patients with severe urethral strictures (mean number of earlier urethrotomies 2.5) were involved. An SR-PLLA spiral stent with a bioabsorption time of 10 to 12 months was inserted into the urethra at the stricture site immediately after optical urethrotomy. Uroflowmetry, measurement of postvoiding residual urine, urine culture, and urethroscopy were performed preoperatively and at 1 and 3 months and then every 3 months up to 12 months, with subsequent follow-up visits every 6 months. The mean follow-up was 46 months. The treatment was successful in 8 of the 22 patients (36%). Six recurrences were found within the stented area and 15 outside. Patients with failure have since been treated with several urethrotomies and repeat self-dilatations or free skin urethroplasties. The results of the use of a bioabsorbable SR-PLLA urethral stent for the treatment of recurrent strictures were encouraging. Without this additional therapy, the recurrence rate of strictures might have been much higher. The main problem was sudden collapse of the stent, possibly induced by outer compression. A new generation of bioabsorbable stents is already under development.
- 1 September 1999
- Published by SAGE Publications in Journal of Endourology
- Vol. 13 (7) , 525-530
Tissue Biocompatibility of a New Caprolactone-Coated Self-Reinforced Self-Expandable Poly-L-Lactic Acid Bioabsorbable Urethral Stent
Background and Purpose: The bioabsorption time as well as other properties of bioabsorbable polymers can be affected by the choice of the basic molecule, by the degree of its polymerization, and by the coating material used in the device. The aim of our present study was to evaluate the biocompatibility of a new caprolactone copolymer-coated, self-reinforced poly-L-lactic acid (SR-PLLA) urethral stent by means of a rabbit muscle implantation test. This new material has previously been tested for cytotoxicity using the thymidine incorporation method (DNA synthesis inhibition test), no toxicity being evidenced. Materials and Methods: Fifteen male rabbits were used as experimental animals. Rods made from pure lactic acid, pure caprolactone copolymer, and caprolactone-coated lactic acid were placed on both sides of the dorsal muscles, eight implants per rabbit. Rods made from latex and silicone were used as positive and negative controls, respectively. The animals were sacrificed after 1 week, 1 mont...
- 1 September 1999
- Published by SAGE Publications in Journal of Endourology
- Vol. 13 (7) , 525-530
Tissue Biocompatibility of a New Caprolactone-Coated Self-Reinforced Self-Expandable Poly-L-Lactic Acid Bioabsorbable Urethral Stent
Background and Purpose: The bioabsorption time as well as other properties of bioabsorbable polymers can be affected by the choice of the basic molecule, by the degree of its polymerization, and by the coating material used in the device. The aim of our present study was to evaluate the biocompatibility of a new caprolactone copolymer-coated, self-reinforced poly-L-lactic acid (SR-PLLA) urethral stent by means of a rabbit muscle implantation test. This new material has previously been tested for cytotoxicity using the thymidine incorporation method (DNA synthesis inhibition test), no toxicity being evidenced. Materials and Methods: Fifteen male rabbits were used as experimental animals. Rods made from pure lactic acid, pure caprolactone copolymer, and caprolactone-coated lactic acid were placed on both sides of the dorsal muscles, eight implants per rabbit. Rods made from latex and silicone were used as positive and negative controls, respectively. The animals were sacrificed after 1 week, 1 mont...