Algorithm Optimization for Multitined Radiofrequency Ablation: Comparative Study in ex Vivo and in Vivo Bovine Liver
Authors:
- Liat AppelbaumLALiat Appelbaum
- Jacob SosnaJSJacob Sosna
- Robert PearsonRPRobert Pearson
- Sarah PerezSPSarah Perez
- Yizhak NissenbaumYNYizhak Nissenbaum
- Pawel MertynaPMPawel Mertyna
- Eugene LibsonELEugene Libson
- S. Nahum GoldbergSGS. Nahum Goldberg
- 1 February 2010
- journal article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 254 (2) , 430-440
Abstract
To prospectively optimize multistep algorithms for largest available multitined radiofrequency (RF) electrode system in ex vivo and in vivo tissues, to determine best energy parameters to achieve large predictable target sizes of coagulation, and to compare these algorithms with manufacturer's recommended algorithms.Institutional animal care and use committee approval was obtained for the in vivo portion of this study. Ablation (n = 473) was performed in ex vivo bovine liver; final tine extension was 5-7 cm. Variables in stepped-deployment RF algorithm were interrogated and included initial current ramping to 105 degrees C (1 degrees C/0.5-5.0 sec), the number of sequential tine extensions (2-7 cm), and duration of application (4-12 minutes) for final two to three tine extensions. Optimal parameters to achieve 5-7 cm of coagulation were compared with recommended algorithms. Optimal settings for 5- and 6-cm final tine extensions were confirmed in in vivo perfused bovine liver (n = 14). Multivariate analysis of variance and/or paired t tests were used.Mean RF ablation zones of 5.1 cm +/- 0.2 (standard deviation), 6.3 cm +/- 0.4, and 7 cm +/- 0.3 were achieved with 5-, 6-, and 7-cm final tine extensions in a mean of 19.5 min +/- 0.5, 27.9 min +/- 6, and 37.1 min +/- 2.3, respectively, at optimal settings. With these algorithms, size of ablation at 6- and 7-cm tine extension significantly increased from mean of 5.4 cm +/- 0.4 and 6.1 cm +/- 0.6 (manufacturer's algorithms) (P <.05, both comparisons); two recommended tine extensions were eliminated. In vivo confirmation produced mean diameter in specified time: 5.5 cm +/- 0.4 in 18.5 min +/- 0.5 (5-cm extensions) and 5.7 cm +/- 0.2 in 21.2 min +/- 0.6 (6-cm extensions).Large zones of coagulation of 5-7 cm can be created with optimized RF algorithms that help reduce number of tine extensions compared with manufacturer's recommendations. Such algorithms are likely to facilitate the utility of these devices for RF ablation of focal tumors in clinical practice.Keywords
This publication has 36 references indexed in Scilit:
- Radiofrequency ablation of hepatic tumors: Lessons learned from 3000 proceduresJournal of Gastroenterology and Hepatology, 2008
- Ablative techniques for the management of kidney cancerNature Reviews Endocrinology, 2007
- Current role of image-guided ablative therapies in lung cancerExpert Review of Anticancer Therapy, 2005
- CT Imaging Findings of Pulmonary Neoplasms After Treatment with Radiofrequency Ablation: Results in 32 TumorsAmerican Journal of Roentgenology, 2005
- Ablative therapies for the treatment of malignant diseases of the breastThe American Journal of Surgery, 2005
- Early-Stage Hepatocellular Carcinoma in Patients with Cirrhosis: Long-term Results of Percutaneous Image-guided Radiofrequency AblationRadiology, 2005
- Ablative Approaches to the Minimally Invasive Treatment of Breast CancerThe Cancer Journal, 2005
- Percutaneous Image-Guided Radiofrequency Ablation of Painful Metastases Involving Bone: A Multicenter StudyJournal of Clinical Oncology, 2004
- Percutaneous Radio-frequency Ablation of Hepatic Metastases from Colorectal Cancer: Long-term Results in 117 PatientsRadiology, 2001
- Percutaneous Radiofrequency Coagulation of Osteoid Osteoma Compared with Operative Treatment*Journal of Bone and Joint Surgery, 1998