Decentration measurements using Placido corneal tangential curvature topography and Scheimpflug tomography pachymetry difference maps after small-incision lenticule extraction
- 1 August 2019
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 45 (8) , 1067-1073
Abstract
To assess the agreement of decentration measurements between tangential curvature topography and tomography pachymetry difference maps after a small-incision lenticule extraction (SMILE) procedure. Yonsei University College of Medicine and Eyereum Eye Clinic, Seoul, South Korea. Retrospective case series. Decentrations and pupillary offset (angle κ) were analyzed and compared using two different methods. A difference map of the tangential curvature was generated with a Placido corneal topographer (Keratron Scout) and a pachymetry difference map was generated with a Scheimpflug tomographer (Pentacam) for each eye, using preoperative and 3-month postoperative scans. Decentrations were also compared using a vector difference plot for the two maps. The study comprised 155 eyes (155 patients). The magnitude of decentration was 0.27 mm ± 0.17 (SD) (range 0.02 to 0.88 mm) with the tangential curvature difference maps and 0.26 ± 0.20 mm (range 0.00 to 0.91 mm) with the pachymetry difference maps, showing no significant difference in the arithmetic values of decentration (P = .124). However, the vector difference in pupillary offset between preoperative and postoperative measurements from the pachymetry difference map significantly correlated with the magnitude of vector differences of decentration between the two maps (P = .007). The vector difference plot also showed a wide distribution spread to all 4 quadrants, demonstrating a lack of agreement. Although there was no significant difference between the scalar magnitude of decentration with the tangential topography difference maps and tomography pachymetry difference maps, the vector differences did not agree and increasing error was present with increasing differences in pupillary offset in the pachymetry difference maps.Keywords
Funding Information
- Korea Health Industry Development Institute
- Ministry of Health & Welfare (HI18C1111)
- Catholic Kwandong University International St. Mary's Hospital (CKU-201706050001)
This publication has 14 references indexed in Scilit:
- Optical Zone Centration Accuracy Using Corneal Fixation-based SMILE Compared to Eye Tracker-based Femtosecond Laser-assisted LASIK for MyopiaJournal of Refractive Surgery, 2015
- LASIK Ablation Centration: An Objective Digitized Assessment and Comparison Between Two Generations of an Excimer LaserJournal of Refractive Surgery, 2015
- Error in the Estimation of Ablation Centration Using Pachymetric Difference MapsJournal of Refractive Surgery, 2015
- Topographic Analysis of the Centration of the Treatment Zone After SMILE for Myopia and Comparison to FS-LASIK: Subjective Versus Objective AlignmentJournal of Refractive Surgery, 2014
- Mild Decentration Measured by a Scheimpflug Camera and Its Impact on Visual Quality Following SMILE in the Early Learning CurveInvestigative Ophthalmology & Visual Science, 2014
- Corneal pachymetry: New ways to look at an old measurementJournal of Cataract & Refractive Surgery, 2014
- Tangential Topography Corneal Map to Diagnose Laser Treatment DecentrationJournal of Refractive Surgery, 2007
- Maximum permissible lateral decentration in aberration-sensing and wavefront-guided corneal ablationJournal of Cataract & Refractive Surgery, 2003
- A Review of Methods for Measuring Corneal TopographyOptometry and Vision Science, 2001
- Corneal topography: A review of terms and conceptsJournal of Cataract & Refractive Surgery, 1996