Head and Neck Imaging with a Dental CBCT Device: Comparison of 360° and 180° Rotation Angles in Effective Dose and Quantitative Image Quality in a Phantom Study
Authors:
Open Access
- 29 May 2025
- journal article
- Published by Bentham Science Publishers Ltd. in Current Medical Imaging
- Vol. 21, 1-13
Abstract
Objectives: This study aims to investigate the effect of full- and half-rotation angles on patient radiation dose and quantitative image quality in CBCT imaging of the head and neck region. Methods: A total of 67 TLDs were used for the dosimetry of 16 different regions in the head and neck slices of the anthropometric phantom. The Hyperion X9 Pro (MyRay, Cefla, Imola, Italy) CBCT device was used with a 90 kV pulsed beam and a 13x16e FOV size. Two separate imaging modes (Regular 360 0 and Quick 180 0) were tested, and the mA was determined by the software. Effective doses (EDs) were calculated using the coefficients recommended by ICRP 103 (2007). For the quantitative image quality tests, three VOIs were manually selected for three separate densities in image slices selected from the mandible, maxilla, and paranasal sinus regions of both volumes separately. Pixel values were averaged, and (SNR), contrast-to-noise ratio (CNR), and uniformity tests were conducted. Results: In 360 0, ED was calculated as 1.903 mSv and the highest absorbed dose was found in the oral mucosa (1.566 mSv). In 180 0, ED was calculated as 1.123 mSv and the highest absorbed dose was found in the right temporal squamous region (0.984 mSv). The reduction in ED was found to be 41% for full- and half-rotation angles. Quick/Regular ratios for SNR and CNR were changed between 0.83-0.91. Conclusion: The magnitude of reduction in ED was found to be higher than the quantitative image quality; however, the impact of this change on diagnosis should be analyzed according to the imaging purpose.Keywords
This publication has 31 references indexed in Scilit:
- Low-dose CBCT protocols in implant dentistry: a systematic reviewOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2024
- Precision of orthodontic cephalometric measurements on ultra low dose-low dose CBCT reconstructed cephalogramsClinical Oral Investigations, 2021
- Metal artifact reduction using iterative CBCT reconstruction algorithm for head and neck radiation therapy: A phantom and clinical studyEuropean Journal of Radiology, 2020
- Optimization of image quality and radiation dose using different cone-beam CT exposure parametersEuropean Journal of Radiology, 2019
- Novel low-dose protocols using cone beam computed tomography in dental medicine: a review focusing on indications, limitations, and future possibilitiesClinical Oral Investigations, 2019
- Factors influencing the effective dose associated with CBCT: a systematic reviewClinical Oral Investigations, 2018
- Cone Beam Computed TomographyDental Clinics of North America, 2018
- Radiation dose in non-dental cone beam CT applications: a systematic reviewLa radiologia medica, 2018
- Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applicationsEuropean Journal of Radiology, 2009
- A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary resultsEuropean Radiology, 1998