A Passively Controlled Biventricular Support Device
- 4 June 2010
- journal article
- Published by Wiley in Artificial Organs
- Vol. 34 (6) , 473-480
Abstract
Clinical studies have reported the balancing of pump outputs to be a serious control issue for rotary biventricular support (BiVS) systems. Poor reliability of long-term, blood immersed pressure sensors encouraged the development of a new control strategy to improve their viability. A rotary BiVS device was designed and constructed with a mechanical passive controller to autoregulate pump outputs to emulate the native baroreceptor response. In vitro testing in a dual circuit, hydraulic mock circulation loop showed that the prototype was able to maintain arterial pressures when subjected to sudden induced hemodynamic destabilization. However, inlet suction was observed when sudden simulated hypertension briefly reduced venous return to the cannulated ventricle. The results have encouraged further development of the device as a means to create an inherently stable, fully passive biventricular support device.Keywords
This publication has 11 references indexed in Scilit:
- The BiVACOR Rotary Biventricular Assist Device: Concept and In Vitro InvestigationArtificial Organs, 2008
- Third‐generation Blood Pumps With Mechanical Noncontact Magnetic BearingsArtificial Organs, 2006
- Progress of Rotary Blood PumpsArtificial Organs, 2006
- Physiological Control of Blood Pumps Using Intrinsic Pump Parameters: A Computer Simulation StudyArtificial Organs, 2006
- A Complete Mock Circulation Loop for the Evaluation of Left, Right, and Biventricular Assist DevicesArtificial Organs, 2005
- Rotary Blood Pumps for Cardiac Assistance: A “Must”?Artificial Organs, 2004
- Assessing the Calf Pulmonary Function During a Long‐Term Biventricular Assist Device Study with a Centrifugal Blood PumpArtificial Organs, 2002
- The Index of Motor Current Amplitude Has Feasibility in Control for Continuous Flow Pumps and Evaluation of Left Ventricular FunctionArtificial Organs, 2001
- Sensorless Controlling Method for a Continuous Flow Left Ventricular Assist DeviceArtificial Organs, 2000
- Detection of Suction and Regurgitation of the Implantable Centrifugal Pump Based on the Motor Current Waveform Analysis and Its Application to Optimization of Pump FlowArtificial Organs, 1999