Tidal volume transmission during non-synchronized nasal intermittent positive pressure ventilation via RAM® cannula
- 12 February 2019
- journal article
- research article
- Published by Springer Nature in Journal of Perinatology
- Vol. 39 (5) , 723-729
Abstract
Nasal intermittent positive pressure ventilation (NIPPV) is a widely used mode of support in neonates, during which ventilator inflations may or may not coincide with spontaneous breathing. We tested the hypothesis that inflations delivered with NIPPV via RAM® cannula and not accompanied by patient effort produce minimal tidal volume as measured by respiratory inductance plethysmography. Fourteen subjects were monitored while receiving NIPPV. We compared tidal volumes during ventilator-supported breaths, unsupported breaths, and ventilator inflations not accompanied by patient effort (defined using electrical activity of the diaphragm). Mean tidal volumes in arbitrary units were 0.30 ± 0.22 in NIPPV inflations associated with patient effort and 0.27 ± 0.15 in spontaneous breaths without ventilator assistance (p = 0.82). Tidal volumes during ventilator-only inflations were 0.06 ± 0.04 (p < 0.005 vs. both ventilator-assisted and unassisted efforts). NIPPV via RAM cannula produces minimal, clinically insignificant tidal volumes during non-spontaneous inflations.Keywords
Funding Information
- Arkansas Children's Hospital Research Institute
This publication has 10 references indexed in Scilit:
- Noninvasive Respiratory SupportPediatrics, 2016
- Evaluating the Effect of Flow and Interface Type on Pressures Delivered With Bubble CPAP in a Simulated ModelRespiratory Care, 2015
- Neonatal nasal intermittent positive pressure ventilation efficacy and lung pressure transmissionJournal of Perinatology, 2015
- Evaluation of a Nasal Cannula in Noninvasive Ventilation Using a Lung SimulatorRespiratory Care, 2014
- Sources of Methodological Variability in Phase Angles from Respiratory Inductance Plethysmography in Preterm InfantsAnnals of the American Thoracic Society, 2014
- A Trial Comparing Noninvasive Ventilation Strategies in Preterm InfantsNew England Journal of Medicine, 2013
- Pressure variation during ventilator generated nasal intermittent positive pressure ventilation in preterm infantsArchives of Disease in Childhood - Fetal and Neonatal Edition, 2010
- Synchronized Nasal Intermittent Positive-Pressure Ventilation and Neonatal OutcomesPediatrics, 2009
- Pulmonary Complications of Mechanical Ventilation in NeonatesClinics in Perinatology, 2008
- Historical Perspectives: Forty Years of Mechanical Ventilation ...Then and Now...NeoReviews, 2003