Development, validation and comparison of two stability-indicating RP-LC methods using charged aerosol and UV detectors for analysis of lisdexamfetamine dimesylate in capsules
Authors:
- Graciela CarlosGCGraciela Carlos
- Eloisa ComiranECEloisa Comiran
- Marcella Herbstrith de OliveiraMOMarcella Herbstrith de Oliveira
- Renata Pereira LimbergerRLRenata Pereira Limberger
- Ana Maria BergoldABAna Maria Bergold
- Pedro Eduardo FröehlichPFPedro Eduardo Fröehlich
Open Access
- 1 November 2016
- journal article
- Published by Scientific Scholar in Arabian Journal of Chemistry
- Vol. 9, S1905-S1914
Abstract
Two new stability-indicating liquid chromatographic methods using two detectors, an ultraviolet (UV) and a charged aerosol detector (CAD) simultaneously connected in series were validated for the assessment of lisdexamfetamine dimesylate (LDX) in capsule. The method was optimized and the influence of individual parameters on UV and CAD response and sensitivity was studied. Chromatography was performed on a Zorbax CN column (250mm×4.6mm, 5μm) in an isocratic elution mode, using acetonitrile and 20mM ammonium formate at pH 4.0 (50:50, v/v) as mobile phase and UV detection at 207nm. The developed method was validated according to ICH guidelines and the parameters’ specificity, limit of detection, limit of quantitation, linearity, accuracy, precision and robustness were evaluated. CAD is designated to be a non-linear detector in a wide dynamic range, however, the method was linear over the concentration range of 70–130μgmL−1 in both detectors. The method was precise and accurate. Robustness study was performed by a Plackett–Burman design, delivering results within the acceptable range. Neither the excipients nor the degradation products showed interference in the method after studies of specificity as well as under stress conditions. The results of the LC-UV and LC-CAD methods were statistically compared through ANOVA and showed no significant difference (p>0.05). Both proposed methods could be considered interchangeable and stability-indicating, and can be applied as an appropriate quality control tool for routine analysis of LDX in capsuleKeywords
This publication has 29 references indexed in Scilit:
- A comparative study on the analytical performance of a charged aerosol detector and an ultraviolet detector for the RP-LC analysis of dabigatran etexilate in capsulesAnalytical Methods, 2013
- Charged aerosol detection in pharmaceutical analysisJournal of Pharmaceutical and Biomedical Analysis, 2012
- Analysis of volatile bases by high performance liquid chromatography with aerosol-based detectionJournal of Chromatography A, 2012
- P.7.c.002 Low variability in exposure to d-amphetamine in children and adults after administration of lisdexamfetamine dimesylateEuropean Neuropsychopharmacology, 2011
- Pharmacokinetics of lisdexamfetamine dimesylate and its active metabolite, d-amphetamine, with increasing oral doses of lisdexamfetamine dimesylate in children with attention-deficit/hyperactivity disorder: A single-dose, randomized, open-label, crossover studyClinical Therapeutics, 2010
- Design and Analysis of Method Equivalence StudiesAnalytical Chemistry, 2009
- Lisdexamfetamine Dimesylate and Mixed Amphetamine Salts Extended-Release in Children with ADHD: A Double-Blind, Placebo-Controlled, Crossover Analog Classroom StudyBiological Psychiatry, 2007
- Development and Testing of a Detection Method for Liquid Chromatography Based on Aerosol ChargingAnalytical Chemistry, 2002
- Evaporative light‐scattering detection coupled to microcolumn liquid chromatography for the analysis of underivatized amino acids: Sensitivity, linearity of response and comparisons with UV absorbance detectionJournal of Microcolumn Separations, 2001
- Clinical Practice Guideline: Diagnosis and Evaluation of the Child With Attention-Deficit/Hyperactivity DisorderPediatrics, 2000