In vivo Assessment of a Manual Single Lumen Alternating Micro-Batch Hemodiafiltration System
Authors:
- Sabrina L. LankerSLSabrina L. Lanker
- Christopher J. PinoCPChristopher J. Pino
- H. David HumesHHH. David Humes
- Lakhmir ChawlaLCLakhmir Chawla
- Kimberly A. JohnstonKJKimberly A. Johnston
- 27 March 2024
- journal article
- research article
- Published by S. Karger AG in Blood Purification
- Vol. 53 (8) , 650-656
Abstract
Introduction: The manual single lumen alternating micro-batch hemodiafiltration (mSLAMB) system is a closed-loop dialysis system designed to provide kidney support in emergency situations (e.g., fluid overload, hyperkalemia, acidemia). If done repeatedly in small batches and at high flow rates, this system was found to achieve clearance levels comparable to traditional renal replacement therapy. Methods: Using a porcine model, uremic toxins and exogenous fluorescent tracer concentrations were successfully lowered within just 1 h of treatment. Results: With a maximal dialysate flow, mSLAMB can achieve decreases in serum potassium concentration of >0.5 mmol/L/h. With the mSLAMB hemodiafiltration system, micro-batch processing was also successful in removing up to 250 mL of ultrafiltrate in 8 cycles. Conclusion: This process could create a better fluid balance allowing for administering therapeutic fluids such as sodium bicarbonate in the clinic. Electrolyte imbalance and volume overload remain severe life-threatening emergencies in low resource settings; therefore, mSLAMB should be explored further due to its modest vascular access requirements, low cost, and ability to be performed without electricity or batteries.Keywords
This publication has 5 references indexed in Scilit:
- Estimating Changes in Glomerular Filtration Rate With Fluorescein Isothiocyanate-Sinistrin During Renal Replacement TherapyASAIO Journal, 2023
- Single Lumen Alternating Micro-Batch Hemodiafiltration (SLAMB-HDF): A Device for Minimally Invasive Renal Replacement TherapyKidney360, 2020
- Peritoneal Dialysis: The Great Savior during DisastersPeritoneal Dialysis International: Journal of the International Society for Peritoneal Dialysis, 2013
- Renal Replacement Therapy in Austere EnvironmentsInternational Journal of Nephrology, 2011
- Fluorescein-labeled sinistrin as marker of glomerular filtration rateEuropean Journal of Medicinal Chemistry, 2005