Hydrogel Expandable Coils for Vascular Occlusion in Congenital Cardiovascular Disease: A Single Center Experience
- 20 October 2011
- journal article
- Published by Tech Science Press in Congenital Heart Disease
- Vol. 7 (3) , 212-218
Abstract
Objective. The objective of this study is to evaluate the safety and feasibility of the AZUR hydrogel-polymer coated platinum coil (hydrocoil) for vascular occlusion in patients with congenital cardiovascular disease (CCVD). Design. Retrospective case review. Patients. Eight patients with CCVD who underwent attempted vascular occlusion procedure with the hydrocoil at the University of Michigan between January 1, 2008 and March 1, 2011. Median patient age was 1.6 years (range 0.6–27.5 years) and median weight was 10.6 kg (3.9–74.8 kg). Outcome Measures. Complete vascular occlusion of the target vessel was assessed at procedural completion with angiography and noninvasive imaging at follow-up. Key procedural variables and adverse events were also evaluated. Results. A total of 21 hydrocoils were successfully placed in seven vessels, in six patients. Vessels included four (57%) veno-venous collaterals, two (29%) aortopulmonary collaterals, and one (14%) main pulmonary artery. Hydrocoils could not be placed successfully in two patients, including one baffle leak and one veno-venous collateral. Pretreatment of the hydrocoil with steam to soften the coil, prior to vascular introduction, increased the likelihood of successful implant (100% vs. 60%). Complete occlusion was observed in 100% of vessels either at the time of catheterization or at follow-up. There were no complications attributable to the hydrocoil. Conclusions. Hydrocoils are safe and effective devices that can be utilized for vascular occlusion in a variety of blood vessels in patients with CCVD. Coil pretreatment may increase the likelihood of procedural success.Keywords
This publication has 12 references indexed in Scilit:
- Embolization of Bleeding Duodenal Ulcer Using Amplatzer Vascular Plug II and Hydrogel Coils: Case ReportVascular and Endovascular Surgery, 2011
- The amplatzer vascular plug and amplatzer vascular plug II for vascular occlusion procedures in 50 patients with congenital cardiovascular diseaseCatheterization and Cardiovascular Interventions, 2010
- HydroCoils, Occlusion Rates, and Outcomes: A Large Single-Center StudyAmerican Journal of Neuroradiology, 2010
- Histomorphology of thrombus organization, neointima formation, and foreign body response in retrieved human aneurysms treated with hydrocoil devicesJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2010
- Percutaneous closure of patent ductus arteriosus: A multiinstitutional registry comparing multiple devicesCatheterization and Cardiovascular Interventions, 2010
- Transcatheter closure of patent ductus arteriosus and aorto-pulmonary vessels using non-ferromagnetic Inconel MReye embolization coilsCatheterization and Cardiovascular Interventions, 2008
- Comparison of two transcatheter device strategies for occlusion of the patent ductus arteriosusCatheterization and Cardiovascular Interventions, 2008
- HydroCoil Endovascular Aneurysm Occlusion and Packing Study (HELPS Trial): Procedural Safety and Operator-Assessed Efficacy ResultsAmerican Journal of Neuroradiology, 2008
- Evaluation of the AMPLATZER vascular plug for embolization of peripheral vascular malformations associated with congenital heart diseaseCatheterization and Cardiovascular Interventions, 2005