Magnetic field interactions in adjustable hydrocephalus shunts
- 1 September 2008
- journal article
- research article
- Published by Journal of Neurosurgery Publishing Group (JNSPG) in Journal of neurosurgery. Pediatrics
- Vol. 2 (3) , 222-228
Abstract
Object: Exposing patients with ventricular shunts to magnetic fields and MR imaging procedures poses a significant risk of unintentional changes in shunt settings. Shunt valves can also generate considerable imaging artifacts. The purpose of this study was to determine the magnetic field safety and MR imaging compatibility of 5 adjustable models of hydrocephalus shunts. Methods: The Codman Hakim (regular and with SiphonGuard), Miethke ProGAV, Medtronic Strata, Sophysa Sophy and Polaris programmable valves were tested in a low-intensity magnetic field, and then translational attraction (TA), magnetic torque (MT), and volume of artifacts on T1-weighted spin echo (SE) and gradient echo (GE) pulse sequences in a 3-T MR imaging unit were measured. Results: The ProGAV and Polaris valves were immune to unintentional reprogramming by magnetic fields up to 3 T. Other valves randomly changed settings, starting from the intensity of field: Sophy valve 24 mT, Strata valve 30 mT, and both Codman Hakim programmable valves from 42 mT. Shunt performances in the 3-T MR imaging unit are reported in the order of compatibility: 1) Codman Hakim regular, TA = 0.005 N, MT = 0.000 Nm, GE = 30 cm3, SE = 2 cm3; 2) Miethke ProGAV, TA = 0.001 N, MT = 1.4 × 10−3 Nm, GE = 231 cm3, SE = 13 cm3; 3) Codman Hakim with SiphonGuard, TA = 0.005 N, MT = 2.3 × 10−3 Nm, GE = 233 cm3, SE = 19 cm3; 4) Medtronic Strata, TA = 0.27 N, MT = 18.0 × 10−3 Nm, GE = 484 cm3, SE = 86 cm3; 5) Sophysa Sophy, TA = 0.82 N, MT = 38.9 × 10−3 Nm, GE = 758 cm3, SE = 72 cm3; and 6) Sophysa Polaris, TA = 0.80 N, MT = 39.6 × 10−3 Nm, GE = 954 cm3, SE = 100 cm3. Conclusions: All valves, with the exception of the Polaris and ProGAV models, are prone to unintentional reprogramming when exposed to heterogeneous magnetic fields stronger than 40 mT. All tested valves can be considered safe for 3-T MR imaging. All valves generated a distortion of the MR image, especially the GE sequences.Keywords
This publication has 10 references indexed in Scilit:
- Experience with the Strata valve in the management of shunt overdrainageJournal of neurosurgery. Pediatrics, 2007
- Adjustable vs set-pressure valves decrease the risk of proximal shunt obstruction in the treatment of pediatric hydrocephalusChild's Nervous System, 2006
- Programmable shunt-related suicide attemptActa Neurochirurgica, 2006
- In vitro hydrodynamic properties of the Miethke proGAV hydrocephalus shuntFluids and Barriers of the CNS, 2006
- Early programmable valve malfunctions in pediatric hydrocephalusJournal of Neurosurgery, 2005
- Effect of 3T MRI on the function of shunt valves—Evaluation of Paedi GAV, Dual Switch and proGAVEuropean Journal of Radiology, 2005
- Evaluation of three new models of hydrocephalus shunts2004
- Effect of Vagus Nerve Stimulator Magnet on Programmable Shunt SettingsNeurosurgery, 2004
- Laboratory Testing of Hydrocephalus Shunts – Conclusion of the U.K. Shunt Evaluation ProgrammeActa Neurochirurgica, 2002
- Electromagnetic field hazards involving adjustable shunt valves in hydrocephalusJournal of Neurosurgery, 2002