In vitro hydrodynamic properties of the Miethke proGAV hydrocephalus shunt
Open Access
- 29 June 2006
- journal article
- Published by Springer Nature in Fluids and Barriers of the CNS
- Vol. 3 (1) , 9
Abstract
Background: Adjustable shunts are very popular in the management of hydrocephalus and are believed to help in minimizing the number of surgical revisions. The drawback with almost all constructions is that they may be accidentally readjusted in relatively weak magnetic fields (around 30–40 mTesla) Materials and methods: The ProGav Miethke shunt is composed of an adjustable ballon-spring valve unit and an integrated over-drainage compensating gravitational device (known as the shunt assistant). A mechanical 'brake' is intended to prevent changes to the valve's performance level in a strong magnetic field. We evaluated the performance and hydrodynamic properties of a sample of three valves in the UK Shunt Evaluation Laboratory. Results: All the shunts showed good mechanical durability over the three-month period of testing, and good stability of hydrodynamic performance over a one-month period The pressure-flow performance curves, operating, opening and closing pressures fell within the limits specified by the manufacturer, and changed according to the programmed performance levels. The operating pressure increased when the shunt assistant was in the vertical position, as specified. The valve has a low hydrodynamic resistance (0.53 mm mmHg ml-1 min-1). External programming proved to be easy and reliable. Strong magnetic fields from a 3 Tesla MR scanner were not able to change the programming of the valve. Conclusion: The ProGAV shunt is an adjustable, low resistance valve that is able to limit posture-related over-drainage. Unlike other adjustable valves, the ProGAV cannot be accidentally re-adjusted by external magnetic field such as a 3T MR scanner.Keywords
This publication has 10 references indexed in Scilit:
- Effect of 3T MRI on the function of shunt valves—Evaluation of Paedi GAV, Dual Switch and proGAVEuropean Journal of Radiology, 2005
- Intraventricular or lumbar infusion test in adult communicating hydrocephalus? Practical consequences and clinical outcome of shunt operationActa Neurochirurgica, 2005
- Effect of 3-tesla magnetic resonance imaging on various pressure programmable shunt valvesJournal of Neurosurgery, 2005
- Effect of cell phone magnetic fields on adjustable cerebrospinal fluid shunt valvesSurgical Neurology, 2005
- Hydrocephalus shunts and waves of intracranial pressureMedical & Biological Engineering & Computing, 2005
- The enigma of underdrainage in shunting with hydrostatic valves and possible solutions2004
- Adjustment and malfunction of a programmable valve after exposure to toy magnetsJournal of neurosurgery. Pediatrics, 2004
- Laboratory Testing of Hydrocephalus Shunts – Conclusion of the U.K. Shunt Evaluation ProgrammeActa Neurochirurgica, 2002
- Electromagnetic field hazards involving adjustable shunt valves in hydrocephalusJournal of Neurosurgery, 2002
- Intracranial pressure and cerebrospinal fluid outflow conductance in healthy subjectsJournal of Neurosurgery, 1991