Postimplant Left Ventricular Assist Device Fit Analysis Using Three-Dimensional Reconstruction
- 1 November 2013
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 59 (6) , 586-592
Abstract
Left ventricular assist devices (LVADs) are blood pumps that augment the function of the failing heart to improve perfusion, resulting in improved survival. For LVADs to effectively unload the left ventricle, the inflow cannula (IC) should be unobstructed and ideally aligned with the heart’s mitral valve (MV). We examined IC orientation deviation from a hypothesized conventional angle (45° right-posterior) and the approximate angle for direct IC-MV alignment in many patients. Three-dimensional anatomic models were created from computed tomography scans for 24 LVAD-implanted patients, and angles were measured between the IC and the apical z-axis in both the coronal and the sagittal planes. Common surgical IC angulation was found to be 22 ± 15° rightward and 21 ± 12° posterior from the apical z-axis; 38% (n = 9) of patients fell in this range. Direct IC-MV angulation was found to be 34 ± 8° rightward and 15 ± 7° posterior; only 8% (n = 2) of patients fell in this range. Rightward deviation toward ventricular septal wall and anterior deviation toward LV anterior freewall are associated with mortalities more so than leftward and posterior deviation. In conclusion, anatomic reconstruction may be a useful preoperative tool to obtain general population and patient-specific alignment for optimal LVAD implantation.Keywords
This publication has 24 references indexed in Scilit:
- Flow Analysis of Ventricular Assist Device Inflow and Outflow Cannula Positioning Using a Naturally Shaped Ventricle and Aortic BranchArtificial Organs, 2010
- Misplacement of LVAD inflow cannula leads to insufficient output and tissue hypoperfusionJournal of Artificial Organs, 2010
- Computer-Assisted 3-Dimensional Anthropometry of the ScaphoidOrthopedics, 2010
- Advanced Heart Failure Treated with Continuous-Flow Left Ventricular Assist DeviceNew England Journal of Medicine, 2009
- Bridge to Heart Transplantation with Left Ventricular Assist Device Versus Inotropic Agents in Status 1 PatientsJournal of Cardiac Surgery, 2009
- Malpositioned Left Ventricular Assist Device Cannula: Diagnosis and Management with Transesophageal Echocardiography GuidanceAnesthesia & Analgesia, 2007
- The Cleveland Clinic PediPump: Anatomic Modeling and Virtual Fitting Studies in a Lamb ModelASAIO Journal, 2007
- Diagnosis of Biventricular Assist Device Inflow Cannula ObstructionEchocardiography, 2007
- Assessment of Recurrent Heart Failure Associated with Left Ventricular Assist Device DysfunctionThe Journal of Heart and Lung Transplantation, 2005
- Long-Term Use of a Left Ventricular Assist Device for End-Stage Heart FailureNew England Journal of Medicine, 2001