Outcome of Resynchronization Therapy on Superficial and Endocardial Electrophysiological Findings

Authors:

Abstract
Cardiac resynchronization therapy (CRT) has proven efficacious in the treatment of patients with heart failure and dyssynchronous activation. Currently, we select suitable CRT candidates based on the QRS complex duration (QRSd) and morphology with left bundle branch block being the optimal substrate for resynchronization. To improve CRT response rates, recommendations emphasize attention to electrical parameters both before implant and after it. Therefore, we decided to study activation times before and after CRT on the body surface potential maps (BSPM) and to compare thus obtained results with data from electroanatomical mapping using the CARTO system. Total of 21 CRT recipients with symptomatic heart failure (NYHA II-IV), sinus rhythm, and QRSd ≥150 ms and 7 healthy controls were studied. The maximum QRSd and the longest and shortest activation times (ATmax and ATmin) were set in the BSPM maps and their locations on the chest were compared with CARTO derived time interval and site of the latest (LATmax) and earliest (LATmin) ventricular activation. In CRT patients, all these parameters were measured during both spontaneous rhythm and biventricular pacing (BVP) and compared with the findings during the spontaneous sinus rhythm in the healthy controls. QRSd was 169.7±12.1 ms during spontaneous rhythm in the CRT group and 104.3±10.2 ms after CRT (p

This publication has 18 references indexed in Scilit: