Development of an algorithm to plan and simulate a new interventional procedure
Authors:
Open Access
- 6 April 2015
- journal article
- case report
- Published by Oxford University Press (OUP) in Interactive CardioVascular and Thoracic Surgery
- Vol. 21 (1) , 87-95
Abstract
The number of implanted biological valves for treatment of valvular heart disease is growing and a percentage of these patients will eventually undergo a transcatheter valve-in-valve (ViV) procedure. Some of these patients will represent challenging cases. The aim of this study was to develop a feasible algorithm to plan and in vitro simulate a new interventional procedure to improve patient outcome. In addition to standard diagnostic routine, our algorithm includes 3D printing of the annulus, hydrodynamic measurements and high-speed analysis of leaflet kinematics after simulation of the procedure in different prosthesis positions as well as X-ray imaging of the most suitable valve position to create a ‘blueprint’ for the patient procedure. This algorithm was developed for a patient with a degenerated Perceval aortic sutureless prosthesis requiring a ViV procedure. Different ViV procedures were assessed in the algorithm and based on these results the best option for the patient was chosen. The actual procedure went exactly as planned with help of this algorithm. Here we have developed a new technically feasible algorithm simulating important aspects of a novel interventional procedure prior to the actual procedure. This algorithm can be applied to virtually all patients requiring a novel interventional procedure to help identify risks and find optimal parameters for prosthesis selection and placement in order to maximize safety for the patient.Keywords
This publication has 25 references indexed in Scilit:
- One-year outcomes of the Surgical Treatment of Aortic Stenosis With a Next Generation Surgical Aortic Valve (TRITON) trial: A prospective multicenter study of rapid-deployment aortic valve replacement with the EDWARDS INTUITY Valve SystemThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Assessment of a transcatheter heart valve prosthesis with multidetector computed tomography: in vitro and in vivo imaging characteristicsThe International Journal of Cardiovascular Imaging, 2012
- In vitro assessment of prosthesis type and pressure recovery characteristics: Doppler echocardiography overestimation of bileaflet mechanical and bioprosthetic aortic valve gradientsThe Journal of Thoracic and Cardiovascular Surgery, 2012
- Failed Valve-In-Valve Transcatheter Aortic Valve ImplantationJACC: Cardiovascular Interventions, 2012
- Transcatheter Aortic Valve Implantation for Failing Surgical Aortic Bioprosthetic ValveJACC: Cardiovascular Interventions, 2011
- Influence of Aortic Dimensions on the Hemodynamic Performance of Small Aortic Valve Prostheses: Impact on Patient/Prosthesis MismatchThe Thoracic and Cardiovascular Surgeon, 2011
- Isolated aortic valve replacement in North America comprising 108,687 patients in 10 years: Changes in risks, valve types, and outcomes in the Society of Thoracic Surgeons National DatabaseThe Journal of Thoracic and Cardiovascular Surgery, 2009
- Aortic Valve Replacement: Choice Between Mechanical Valves and BioprosthesesJournal of Cardiac Surgery, 2008
- Very Long-Term Survival Implications of Heart Valve Replacement With Tissue Versus Mechanical Prostheses in Adults <60 Years of AgeCirculation, 2007
- Valve-in-a-Valve Concept for Transcatheter Minimally Invasive Repeat Xenograft ImplantationJACC, 2007