Reproducibility and cost of blood pressure self-measurement using the ‘Loaned Self-measurement Equipment Model’
Authors:
- César Calvo-VargasCCCésar Calvo-Vargas
- Victoria Padilla RiosVRVictoria Padilla Rios
- Rogelio Troyo-SanrománRTRogelio Troyo-Sanromán
- Fernando Grover-PaezFGFernando Grover-Paez
- 1 October 2001
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 6 (5) , 225-232
Abstract
To determine the reproducibility and the cost of the Loaned Self-measurement Equipment Model for blood pressure. Seventy-five subjects with non-complicated essential hypertension, having an average age of 56.2 +/- 11 years, were evaluated. The Loaned Self-measurement Equipment Model consists of a set of sphygmomanometers belonging to the clinic, loaned to patients for 3-day periods for them to carry out 27 readings, and then returned. Using a mercury sphygmomanometer, a nurse carried out office blood pressure readings during two visits. A month later, the same procedure was repeated under the same clinical conditions. The agreement between studies was assessed by correlation coefficients, coefficients of variation and standard deviation of the differences (SDD). The office blood pressure was 147 +/- 17/89 +/- 12 mmHg during visit A, and 145 +/- 16/88 +/- 11 mmHg during visit B. The self-measurement blood pressure was 139 +/- 16/84 +/- 10 mmHg during the first visit, and 136 +/- 15/84 +/- 11 mmHg during the second visit. In neither of the cases were the differences statistically significant. The correlation coefficients for the office and self-measurement systolic blood pressure were statistically significant (P = 0.018), while those for the diastolic blood pressure were not statistically significant. The SDD for self-measurement systolic blood pressure was 10.7, compared to 15.0 for the office blood pressure (P < 0.01). The comparison in the case of diastolic blood pressure was not significant. The cost of each loan was $5.85 (US dollars). The Loaned Self-measurement Equipment Model has a good short-term reproducibility, and the cost is reasonable.Keywords
This publication has 30 references indexed in Scilit:
- Prognosis of Isolated Systolic and Isolated Diastolic Hypertension as Assessed by Self-Measurement of Blood Pressure at HomeArchives of internal medicine (1960), 2000
- Home Blood Pressure MonitoringArchives of internal medicine (1960), 2000
- Value of blood pressure self-monitoring as a predictor of progression of diabetic nephropathyJournal Of Hypertension, 1999
- The International Database of Self-Recorded Blood Pressures in normotensive and untreated hypertensive subjectsBlood Pressure Monitoring, 1999
- Determination of accuracies of 10 models of home blood pressure monitors using an oscillometric simulatorBlood Pressure Monitoring, 1999
- Calcium supplementation in patients with essential hypertensionJournal Of Hypertension, 1998
- Effects of Magnesium Supplementation in Hypertensive PatientsHypertension, 1998
- Home blood pressure measurement has a stronger predictive power for mortality than does screening blood pressure measurementJournal Of Hypertension, 1998
- Hypertension Optimal Treatment (HOT) StudyHypertension, 1998
- Home Blood Pressure MonitoringMedical Care, 1992