Validating kQ =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beam
Open Access
- 1 March 2018
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 19 (3) , 351-354
Abstract
AAPM Report 142 recommends and the State of Ohio requires that the imaging dose be quantified in radiotherapy applications. Using the TG51 dose calibration protocol for MV Imaging dose measurement requires knowledge of the kQ parameter for the beam quality and the ionization chamber type under investigation. The %dd(10)x of the Varian TrueBeam 2.5 MV imaging beam falls outside the range of the available data for the calculation of the kQ value. Due to the similarities of the 2.5 MV imaging beam and the 60Co beam, we and others made the assumption that kQ = 1.0 in TG51 calculations. In this study, we used the TG21 and TG51 calibration protocols in conjunction to validate that kQ = 1.0 for the 2.5 MV imaging beam using a PTW 30013 farmer chamber. Standard measurements for TG51 absolute dosimetry QA were performed at 100 cm SSD, 10 cm depth, 10 × 10 field size, delivering 100 Monitor Units to a waterproof Farmer Chamber (PTW TN30013) for both 2.5 and 6 MV. Both the TG21 and TG51 formalisms were used to calculate the dose to water per MU at dmax (Dw/MU) for the 6 MV beam. The calculated outputs were 1.0005 and 1.0004 cGy/MU respectively. The TG21 formalism was then used to calculate (Dw/MU) for the 2.5 MV imaging beam. This value was then used in the TG51 formalism to find kQ for the 2.5 MV imaging beam. A kQ value of 1.00 ± 0.01 was calculated for 2.5 MV using this method.Keywords
This publication has 7 references indexed in Scilit:
- Characterization of a 2.5 MV inline portal imaging beamJournal of Applied Clinical Medical Physics, 2016
- Addendum to the AAPMˈs TG‐51 protocol for clinical reference dosimetry of high‐energy photon beamsMedical Physics, 2014
- Task Group 142 report: Quality assurance of medical acceleratorsa)Medical Physics, 2009
- Calculated absorbed‐dose ratios, TG51/TG21, for most widely used cylindrical and parallel‐plate ion chambers over a range of photon and electron energiesMedical Physics, 2002
- Comparison between TG-51 and TG-21: Calibration of photon and electron beams in water using cylindrical chambersJournal of Applied Clinical Medical Physics, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983