Orthopedic implants affect the electric field induced by switching gradients in MRI
Open Access
- 29 September 2023
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 91 (1) , 398-412
Abstract
Purpose: To investigate whether the risk of peripheral nerve stimulation increases in the presence of bulky metallic prostheses implanted in a patient's body.Methods: A computational tool was used to calculate the electric field (E‐field) induced in a realistic human model due to the action of gradient fields. The calculations were performed both on the original version of the anatomical model and on a version modified through “virtual surgery” to incorporate knee, hip, and shoulder prostheses. Five exam positions within a body gradient coil and one position using a head gradient coil were simulated, subjecting the human model to the readout gradient from an EPI sequence. The induced E‐field in models with and without prostheses was compared, focusing on the nerves and all other tissues (both including and excluding the bones from the analysis).Results: In the nerves, the most pronounced increase in the E‐field (+24%) was observed around the knee implant during an abdominal MRI (Y axis readout). When extending the analysis to encompass all tissues (excluding bones), the greatest amplification (+360%) occurred around the knee implant during pelvic MRI (Z axis readout). Notable increases in E‐field peaks were also identified around the shoulder and hip implants in multiple scenarios.Conclusion: Based on the presented results, further investigations aimed at quantifying the threshold of nerve stimulation in the presence of bulky implants are desirable.Keywords
Funding Information
- European Metrology Programme for Innovation and Research (17IND01 MIMAS)
- European Partnership on Metrology (21NRM05 STASIS)
This publication has 42 references indexed in Scilit:
- Stable FFT-JVIE solvers for fast analysis of highly inhomogeneous dielectric objectsJournal of Computational Physics, 2014
- Magnetic resonance imaging of the postoperative hipJournal of Magnetic Resonance Imaging, 2012
- Reducing the staircasing error in computational dosimetry of low-frequency electromagnetic fieldsPhysics in Medicine & Biology, 2012
- Reducing peripheral nerve stimulation due to gradient switching using an additional uniform field coilMagnetic Resonance in Medicine, 2011
- Magnetic Resonance Imaging of Joint ReplacementsSeminars in Musculoskeletal Radiology, 2006
- Peripheral Nerve Stimulation by Gradient Switching Fields in Magnetic Resonance ImagingIEEE Transactions on Biomedical Engineering, 2004
- Comparison of the threshold for peripheral nerve stimulation during gradient switching in whole body MR systemsJournal of Magnetic Resonance Imaging, 2002
- Peripheral nerve stimulation in a whole‐body echo‐planar imaging systemJournal of Magnetic Resonance Imaging, 1997
- Calculation of electric fields induced near metal implants by magnetic resonance imaging switched-gradient magnetic fieldsMagnetic Resonance Imaging, 1997
- Theoretical evaluation of peripheral nerve stimulation during MRI with an implanted spinal fusion stimulatorMagnetic Resonance Imaging, 1997