Intermittent Pacemaker Malfunction Caused by Continuous Compression of the Lead by the Clavicle (Subclavian Crush Syndrome)
Authors:
Open Access
- 27 May 2020
- journal article
- Published by SMC Media in European Journal of Case Reports in Internal Medicine
- Vol. 7 (8) , 001684
Abstract
Subclavian vein access is still one of the most favoured access options for cardiac implantable electronic device (CIED) implantation. For the physician, the technique is reasonably familiar and easy to carry out. However, this has several potential complications. In this case, we present a late complication of subclavian access. The patient presented with intermittent loss of pacemaker output, which caused him to experience several syncopal events. In the acute setting, we changed the lead polarity and achieved a good outcome. Further management of this situation consisted of removal and replacement of the damaged lead.Keywords
This publication has 11 references indexed in Scilit:
- 2018 EHRA expert consensus statement on lead extraction: recommendations on definitions, endpoints, research trial design, and data collection requirements for clinical scientific studies and registries: endorsed by APHRS/HRS/LAHRSEP Europace, 2018
- The European Lead Extraction ConTRolled (ELECTRa) study: a European Heart Rhythm Association (EHRA) Registry of Transvenous Lead Extraction OutcomesEuropean Heart Journal, 2017
- Venous access and long-term pacemaker lead failure: comparing contrast-guided axillary vein puncture with subclavian puncture and cephalic cutdownEP Europace, 2016
- Pacemaker insertion2015
- Radiographic Predictors of Lead Conductor FractureCirculation: Arrhythmia and Electrophysiology, 2014
- Preferred tools and techniques for implantation of cardiac electronic devices in Europe: results of the European Heart Rhythm Association surveyEP Europace, 2013
- A Comparison of Lead Placement Through the Subclavian Vein Technique With Fluoroscopy-Guided Axillary Vein Technique for Permanent Pacemaker InsertionCanadian Journal of Cardiology, 2012
- Lead Extraction Is Preferred for Lead Revisions and System Upgrades: When Less Is MoreCirculation: Arrhythmia and Electrophysiology, 2010
- HRS/EHRA Expert Consensus on the Monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of Techniques, Indications, Personnel, Frequency and Ethical ConsiderationsDeveloped in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC), the American Heart Association (AHA), the European Society of Cardiology (ESC), the Heart Failure Association of ESC (HFA), and the Heart Failure Society of America (HFSA). Endorsed by the Heart Rhythm Society, the European Heart Rhythm Association (a registered branch of the ESC), the American College of Cardiology, the American Heart AssociationEP Europace, 2008
- Lead Fracture in Cephalic Versus Subclavian Approach with Transvenous Implantable Cardioverter Defibrillator SystemsPacing and Clinical Electrophysiology, 1996