Out-of-field neutron doses in proton therapy: from experimental characterization to a practical calculation tool
Open Access
- 26 May 2026
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 71 (11)
Abstract
Objective: To experimentally characterize out-of-field neutron dose in a pencil beam scanning (PBS) proton therapy treatment room using multiple detectors, to analyze its dependence on proton energy, gantry angle, distance to isocenter, and field size, and to develop a practical calculation tool for neutron dose estimation. Approach: Two extended-range ambient detectors, bubble (BDs) detectors, track-etch, thermoluminescent (TLDs), and three electronic personal dosimeters were used. Measurements were taken at up to 21 points for five proton energies, four gantry angles (ϴ), four field sizes, four measurement lines, and distances of 1-4.7 m from the isocenter. Room symmetry was evaluated. At each point, H*(10) or Hp(10) = a·E p b was assessed, and measurements were also performed for a clinical irradiation, with the resulting doses compared against weighted sums of contributions from individual energy layers. Bland-Altman analyses were applied. Main results: The room was symmetric for ϴ = 270º-90º, while doses at 0º were on average 8% lower than at 180º. Neutron doses for the spot field and 10 × 10 cm 2 field were interchangeable, with 20 × 20 and 30 × 30 cm 2 fields differing by 2-22%. Dose dependence on proton energy followed a power-law, with best fits for ambient detectors and moderate performance for BDs and TLDs. Linear superposition held for ambient and BDs, validating that neutron doses can be approximated as weighted sums of contributions from individual energy layers. A Python-based calculator for out-of-field neutron doses, provided as supplementary material, was developed for six detectors. Significance: This study delivers an easy-to-use Python tool for first-order estimation of out-of-field neutron doses in a PBS proton therapy room for any treatment plan or location. The tool is based on the linear superposition model, is validated against measurements, and accounts for detector-dependent uncertainties.Keywords
Funding Information
- Sociedad Española de Física Médica (PI-SEFM-2023)
- Instituto de Salud Carlos III (PI23/00157)
This publication has 28 references indexed in Scilit:
- Radiation protection in proton therapy: Insights from a multi-institutional survey and experimental measurements within the SINFONIA projectPhysica Medica, 2025
- Intercomparison of personal and ambient neutron detectors used for radiation protection in a synchrotron-based proton therapy facility: experimental and Monte Carlo resultsPhysica Medica, 2025
- Neutron personal dosimetry using polyallyl diglycol carbonate (PADC): Current status, best practices and proposed researchPhysics Open, 2022
- A review of proton therapy – Current status and future directionsPrecision Radiation Oncology, 2022
- A Monte Carlo-based analytic model of neutron dose equivalent for a mevion gantry-mounted passively scattered proton system for craniospinal irradiationMedical Physics, 2020
- Measurement of stray neutron doses inside the treatment room from a proton pencil beam scanning systemPhysica Medica, 2017
- Neutron field characteristics of Ciemat's Neutron Standards LaboratoryApplied Radiation and Isotopes, 2015
- Configuration and validation of an analytical model predicting secondary neutron radiation in proton therapy using Monte Carlo simulations and experimental measurementsPhysica Medica, 2015
- Measurement of Neutron Dose Equivalent and its Dependence on Beam Configuration for a Passive Scattering Proton Delivery SystemInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- Risk of Developing Second Cancer From Neutron Dose in Proton Therapy as Function of Field Characteristics, Organ, and Patient AgeInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008