Development of a thermodynamic control system for the Fontan circulation pulsation device using shape memory alloy fibers
- 18 April 2015
- journal article
- research article
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 18 (3) , 199-205
Abstract
The Fontan procedure is one of the common surgical treatments for circulatory reconstruction in pediatric patients with congenital heart disease. In Fontan circulation, low pulsatility may induce localized lung ischemia and may impair the development of pulmonary peripheral endothelial cells. To promote pulmonary circulation in Fontan circulation, we have been developing a pediatric pulmonary circulatory pulsation device using shape memory alloy fibers attached from the outside of total cavopulmonary connection. In this study, we developed a new thermal control system for the device and examined its functions. We mounted on the device 16 fibers connected in parallel around an ePTFE graft circumferentially. To provide optimized contraction, we designed the new thermal control system. The system consisted of a thermistor, a pressure sensor, and a regulator that was controlled by the adaptive thermodynamic transfer functions. We monitored the parameters and calculated heat transfer function as well as pressure distribution on the graft surface. Then we examined and compared the dynamic contractile pressure and changes in surface temperature. As a result, by the application of the control based on the new feedback system analysis, the circumferential contractile pressure increased by 35 %. The adaptive thermodynamic regulation was useful for the selection of alternative thresholds of the surface temperature of the graft. The system could achieve effective contraction for the pulsatile flow generation by the device.Keywords
This publication has 28 references indexed in Scilit:
- Effects of lack of pulsatility on pulmonary endothelial function in the Fontan circulationThe Journal of Thoracic and Cardiovascular Surgery, 2013
- Pulmonary Capillary Wedge Pressure Augments Right Ventricular Pulsatile LoadingCirculation, 2012
- ACCF/AHA/HFSA 2011 Survey Results: Current Staffing Profile of Heart Failure Programs, Including Programs That Perform Heart Transplant and Mechanical Circulatory Support Device ImplantationJournal of Cardiac Failure, 2011
- Cavopulmonary Assist: (Em)powering the Univentricular Fontan CirculationSeminars in Thoracic and Cardiovascular Surgery: Pediatric Cardiac Surgery Annual, 2011
- Intravascular Mechanical Cavopulmonary Assistance for Patients With Failing Fontan PhysiologyArtificial Organs, 2009
- Performance of Cavopulmonary Palliation at Elevated AltitudeCirculation, 2008
- Mechanical support of total cavopulmonary connection with an axial flow pumpThe Journal of Thoracic and Cardiovascular Surgery, 2005
- Effects of Exercise and Respiration on Blood Flow in Total Cavopulmonary ConnectionCirculation, 2003
- Evaluation of a Pulsatile Pediatric Ventricular Assist Device in an Acute Right Heart Failure ModelThe Annals of Thoracic Surgery, 1997