Ultrasound-Based Robot-Assisted Drilling for Minimally Invasive Pedicle Screw Placement
- 8 April 2024
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Robotics and Bionics
- Vol. 6 (3) , 818-828
Abstract
Minimally invasive pedicle screw placement (MIPSP) is a widely used treatment for spine diseases. When coupled with intraoperative navigation modalities, robots may help improve surgical outcomes and reduce complications. With such a system, the application of pedicle screws has been expanded from needle insertion to the spine surgery. This paper investigates the possibility and feasibility of robot-assisted MIPSP based on ultrasound (US) guidance. The proposed system is non-radiative and fiducial-free, using purely image information to close the registration loop. Then the system automatically positions the drill tip to a planned screw trajectory and executes the drilling operation. Experiments were conducted on both ex-vivo lamb and human cadaver spines. An entry point accuracy of 2.39±1.41 mm, and orientation accuracy of 2.82±1.85∘ was found for 24 drilled trajectories on three lamb spines. On the ex-vivo human spine, the position error averaged 3.08±2.43 mm at the entry point and 4.05±2.62 mm at the stop point across 16 drilling instances. Moreover, a 87.5% success rate was reported by using Gertzbein-Robbins grade. The experimental results demonstrate the potential for offering a radiation-free alternative. Although restricted to cadaver trials, this work encourages further exploration of this technology to assist surgeons in maximizing performance in clinical practice.Keywords
All Related Versions
Funding Information
- European Union’s Horizon 2020 Research and Innovation Program (101016985 (FAROS))
- Flemish Research Foundation (G0A1420N, 1S36324N, 1S35424N, 1SHGA24N)
This publication has 45 references indexed in Scilit:
- Development and evaluation of robot‐assisted ultrasound navigation system for pedicle screw placement: An ex‐vivo animal validationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2023
- Robot-assisted ultrasound reconstruction for spine surgery: from bench-top to pre-clinical studyInternational Journal of Computer Assisted Radiology and Surgery, 2023
- Clinical application of spinal robot in cervical spine surgery: safety and accuracy of posterior pedicle screw placement in comparison with conventional freehand methodsNeurosurgical Review, 2023
- Robotics in spine surgery: systematic review of literatureInternational Orthopaedics, 2022
- Radiological and clinical differences between robotic-assisted pedicle screw fixation with and without real-time optical trackingEuropean Spine Journal, 2020
- Development and Evaluation of CT-to-3D Ultrasound Image Registration Algorithm in Vertebral Phantoms for Spine SurgeryAnnals of Biomedical Engineering, 2020
- U-Net: Convolutional Networks for Biomedical Image Segmentation2015
- Methods to determine pedicle screw placement accuracy in spine surgery: a systematic reviewEuropean Spine Journal, 2015
- Robot-assisted spine surgery: feasibility study through a prospective case-matched analysisEuropean Spine Journal, 2015
- Accuracy of pedicle screw placement: a systematic review of prospective in vivo studies comparing free hand, fluoroscopy guidance and navigation techniquesEuropean Spine Journal, 2011