Evaluation of a Ventricular Assist Device System
Authors:
- Michael S. GossmanMGMichael S. Gossman
- Joel D. GrahamJGJoel D. Graham
- Indra J. DasIDIndra J. Das
- Chee-Wai ChengCCChee-Wai Cheng
- Len CoutinhoLCLen Coutinho
- Dan TamezDTDan Tamez
- Jeffrey A. LaroseJLJeffrey A. Larose
- 1 November 2012
- journal article
- research article
- Published by Wolters Kluwer Health in ASAIO Journal
- Vol. 58 (6) , 597-600
Abstract
Inadequate research exists regarding testing of a ventricular assist device (VAD) for susceptibility to radiation damage. Specifically, minimal data are available to radiation oncologists prescribing treatment plans for patients with an implanted VAD. As the number of implanted devices increases, patients requiring radiation at tissue sites near or at the device will increase. The purpose of this study is to provide the first analysis of radiation effects of proton beams on VADs. Five left VAD (LVAD) pumps (HeartWare Inc., Miami Lakes, FL) were exposed to proton beam radiation at a calibrated dose rate of 5 Gy/min up to a cumulative dose of 70 Gy. The Heartware LVAD pump recorded parameters including power (W), speed (revolutions/min), and estimated flow (L/min). Analysis of collected data after each irradiation found no deviation in pump parameters from baseline values. The Heartware LVAD pump exhibited no change in device function when directly irradiated by a high energy proton beam. Secondary neutron fluence created in the proton beam during irradiation had no effect on external components including the system controller and batteries powering the Heartware LVAD.Keywords
This publication has 20 references indexed in Scilit:
- Evaluation of a Ventricular Assist DeviceASAIO Journal, 2012
- Results of Proton Beam Therapy without Concurrent Chemotherapy for Patients with Unresectable Stage III Non-small Cell Lung CancerJournal of Thoracic Oncology, 2012
- Third INTERMACS Annual Report: The evolution of destination therapy in the United StatesThe Journal of Heart and Lung Transplantation, 2011
- Design Concepts and Principle of Operation of the HeartWare Ventricular Assist SystemASAIO Journal, 2010
- High-dose radiotherapy exposure to cardiac pacemakers may be safe in selected patientsRadiotherapy and Oncology, 2010
- Continuous Flow Left Ventricular Assist Device Improves Functional Capacity and Quality of Life of Advanced Heart Failure PatientsJACC, 2010
- Extended Mechanical Circulatory Support With a Continuous-Flow Rotary Left Ventricular Assist DeviceJACC, 2009
- Implanted Cardiac Defibrillator Care in Radiation Oncology Patient PopulationInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillatorsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Influence of radiotherapy on the latest generation of pacemakersRadiotherapy and Oncology, 2005