Performance evaluation of the Abbott CELL‐DYN Ruby and the Sysmex XT‐2000i haematology analysers
- 10 January 2011
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 33 (1) , 19-29
Abstract
Summary: Two mid‐range haematology analysers (Abbott CELL‐DYN Ruby and Sysmex XT‐2000i) were evaluated to determine their analytical performance and workflow efficiency in the haematology laboratory. In total 418 samples were processed for determining equivalence of complete blood count (CBC) measurements, and 100 for reticulocyte comparison. Blood smears served for assessing the agreement of the differential counts. Inter‐instrument agreement for most parameters was good although small numbers of discrepancies were observed. Systematic biases were found for mean cell volume, reticulocytes, platelets and mean platelet volume. CELL‐DYN Ruby WBC differentials were obtained with all samples while the XT‐2000i suppressed differentials partially or completely in 13 samples (3.1%). WBC subpopulation counts were otherwise in good agreement with no major outliers. Following first‐pass CBC/differential analysis, 88 (21%) of XT‐2000i samples required further analyser processing compared to 18 (4.3%) for the CELL‐DYN Ruby. Smear referrals for suspected WBC/nucleated red blood cells and platelet abnormalities were indicated for 106 (25.4%) and 95 (22.7%) of the XT‐2000i and CELL‐DYN Ruby samples respectively. Flagging efficiencies for both analysers were found to be similar. The Sysmex XT‐2000i and Abbott CELL‐DYN Ruby analysers have broadly comparable analytical performance, but the CELL‐DYN Ruby showed superior first‐pass efficiency.Keywords
This publication has 14 references indexed in Scilit:
- Rule based processing of the CD4000, CD3200 and CD Sapphire analyser output using the Cerner Discern Expert ModuleInternational Journal of Laboratory Hematology, 2009
- Evaluation of the platelet counting by Abbott CELL-DYN®SAPPHIRE™ haematology analyser compared with flow cytometryInternational Journal of Laboratory Hematology, 2009
- Workflow restrictions on hematology analyzers XE-2100 and XS-800 when assaying pediatric samples with limited volumeClinical Chemistry and Laboratory Medicine (CCLM), 2009
- Performance evaluation of Abbott CELL‐DYN Ruby for routine useInternational Journal of Laboratory Hematology, 2008
- Automated Blood Cell CountsAmerican Journal of Clinical Pathology, 2008
- European Multi-Center Evaluation of the Abbott Cell-Dyn Sapphire Hematology AnalyzerLaboratory Hematology, 2006
- Accurate basophils counting: not an easy goal!Clinical and Laboratory Haematology, 2005
- Automated CD61 immunoplatelet analysis of thrombocytopenic samplesBritish Journal of Haematology, 2001
- Sysmex SE-9000 Hematology Analyzer:Performance Evaluation on Leukocyte Differential Counts Using an NCCLS H20-A ProtocolAmerican Journal of Clinical Pathology, 1997
- Light‐scattering polarization measurements as a new parameter in flow cytometryCytometry, 1987