Investigating the clinical advantages of a robotic linac equipped with a multileaf collimator in the treatment of brain and prostate cancer patients
Open Access
- 8 September 2015
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 16 (5) , 284-295
Abstract
The purpose of this study was to evaluate the performance of a commercially available CyberKnife system with a multileaf collimator (CK-MLC) for stereotactic body radiotherapy (SBRT) and standard fractionated intensity-modulated radiotherapy (IMRT) applications. Ten prostate and ten intracranial cases were planned for the CK-MLC. Half of these cases were compared with clinically approved SBRT plans generated for the CyberKnife with circular collimators, and the other half were compared with clinically approved standard fractionated IMRT plans generated for conventional linacs. The plans were compared on target coverage, conformity, homogeneity, dose to organs at risk (OAR), low dose to the surrounding tissue, total monitor units (MU), and treatment time. CK-MLC plans generated for the SBRT cases achieved more homogeneous dose to the target than the CK plans with the circular collimators, for equivalent coverage, conformity, and dose to OARs. Total monitor units were reduced by 40% to 70% and treatment time was reduced by half. The CK-MLC plans generated for the standard fractionated cases achieved prescription isodose lines between 86% and 93%, which was 2%–3% below the plans generated for conventional linacs. Compared to standard IMRT plans, the total MU were up to three times greater for the prostate (whole pelvis) plans and up to 1.4 times greater for the intracranial plans. Average treatment time was 25 min for the whole pelvis plans and 19 min for the intracranial cases. The CK-MLC system provides significant improvements in treatment time and target homogeneity compared to the CK system with circular collimators, while maintaining high conformity and dose sparing to critical organs. Standard fractionated plans for large target volumes were generated that achieved high prescription isodose levels. The CK-MLC system provides more efficient SRS and SBRT treatments and, in select clinical cases, might be a potential alternative for standard fractionated treatments. PACS numbers: 87.56.nk, 87.56.bd
Keywords
This publication has 28 references indexed in Scilit:
- Advances in fiducial‐free image‐guidance for spinal radiosurgery with CyberKnife – a phantom studyJournal of Applied Clinical Medical Physics, 2010
- Patient Motion and Targeting Accuracy in Robotic Spinal Radiosurgery: 260 Single-Fraction Fiducial-Free CasesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2010
- The CyberKnife® Robotic Radiosurgery System in 2010Technology in Cancer Research & Treatment, 2010
- The design, physical properties and clinical utility of an iris collimator for robotic radiosurgeryPhysics in Medicine & Biology, 2009
- Intrafraction Geometric Uncertainties in Frameless Image-Guided RadiosurgeryInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2009
- Intrafractional Motion of the Prostate During Hypofractionated RadiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2008
- Reducing monitor units for robotic radiosurgery by optimized use of multiple collimatorsMedical Physics, 2008
- A fast, accurate, and automatic 2D–3D image registration for image‐guided cranial radiosurgeryMedical Physics, 2008
- Respiration tracking in radiosurgeryMedical Physics, 2004
- Fiducial‐based targeting accuracy for external‐beam radiotherapyMedical Physics, 2002