Bone fracture healing under Ilizarov fixator: Influence of fixator configuration, fracture geometry, and loading
- 14 March 2019
- journal article
- research article
- Published by Wiley in International Journal for Numerical Methods in Biomedical Engineering
- Vol. 35 (6) , e3199
Abstract
[1] This study aims to enhance the understanding of the relationship between Ilizarov fixator configuration and its effects on bone fracture healing. Using Taylor spatial frame (TSF) as an example, the roles of critical parameters (i.e. TSF ring diameter, wire pretension, fracture gap size and axial load) that govern fracture healing during the early stages were investigated by using computational modelling in conjunction with mechanical testing involving an advanced 3D optical measurement system. The computational model was first validated using the mechanical test results and then used to simulate mesenchymal stem cell (MSC) differentiations within different regions of the fracture site under various combinations of TSF ring diameter, wire pretension, fracture gap size and axial load values. Predicted spatially dependent MSC differentiation patterns and the influence of each parameter on differentiations were compared with in vivo results and good agreement was seen between the two. Gap size was identified as the most influential parameter in MSC differentiation and the influence of axial loading and TSF configuration (i.e. ring diameter and wire pretension) on cell differentiation was seen to be gap size dependent. Most changes in cell differentiation were predicted in the external callus (periosteal) which is the crucial region of the callus in the early stages. However, for small gap sizes (e.g. 1 mm) significant changes were predicted in the endosteal callus as well. The study exhibits the potential of computational models in assessing the performance of Ilizarov fixators as well as assisting surgeons in patient specific clinical treatment planning.Keywords
Funding Information
- University of Melbourne
This publication has 54 references indexed in Scilit:
- Fracture healing under healthy and inflammatory conditionsNature Reviews Rheumatology, 2012
- Theoretical and Finite Element Modeling of Fine Kirschner Wires in Ilizarov External FixatorJournal of Medical Devices, 2010
- Comparison of transverse wires and half pins in Taylor Spatial Frame: A biomechanical studyJournal of Orthopaedic Surgery and Research, 2010
- Analytical modelling of Kirschner wires in Ilizarov circular external fixator as pretensioned slender beamsJournal of The Royal Society Interface, 2008
- Solute transport in cartilage undergoing cyclic deformationComputer Methods in Biomechanics and Biomedical Engineering, 2007
- The Mechanics of External FixationHSS Journal®: The Musculoskeletal Journal of Hospital for Special Surgery, 2007
- Determination of ankle external fixation stiffness by expedited interactive finite element analysisJournal of Orthopaedic Research, 2005
- Comparative performance of a flexible fixation implant to a rigid implant in static and repetitive incremental loadingJournal of Materials Processing Technology, 2005
- The effect of micromovement on callus formationJournal of Orthopaedic Science, 2001
- Biophysical stimuli on cells during tissue differentiation at implant interfacesJournal of Biomechanics, 1997